Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Boston Scientific ROTAPRO, Rotational Atherectomy System Console, Model numbers H74939309010 (US) and H74939309030 (EU).

Recall number
Z-0381-2022
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific ROTAPRO, Rotational Atherectomy System Console, Model numbers H74939309010 (US) and H74939309030 (EU).
Status
Terminated
Initiated
2021-11-08
Posted
not on file
Terminated
2023-05-02
Reason
Manufacturing instructions did not include the directions to not retest the pneumatic kit component of the ROTAPRO console upon failure, an internal nonconformity.
Root cause
Employee error
Action
The recalling firm issued letters dated 11/8/2021 on the same date and flagged "Urgent Medical Device Removal - Immediate Action Required." It explains the reason for recall, the risk to health, and provides removal instructions. The instructions say to immediately discontinue use of and segregate the affected product. A Reply Verification Tracking Form is enclosed for immediate completion and return. The customer will be contacted by BSC and provided a Returned Goods Authorization for return of the device.
Quantity
7 devices
Distribution
Worldwide distribution - US Nationwide distribution in the states of FL, NV, NY, OH, PA, and WA. The countries of Germany.
Lot, serial or model codes
US Serial numbers RP005005, RP005013, RP005017, RP005036, RP005039, and RP005040, GTIN 08714729975076; EU Serial number RP005010, GTIN 08714729975151.
510(k) numbers
not on file
PMA numbers
not on file
Product code
MCX
Firm FEI
3002095335
Firm city
Maple Grove
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register