FDA Medical Device Recalls (openFDA)
Owen Mumford USA, Inc.: Unifine Pentips 5mm x 31G pen needles, packaged in 100 ct pen needles. Product Usage: Sterile, non-toxic, non-pyrogenic, single use pen needles for the subcutaneous injection of drugs from pen injectors.
- Recall number
- Z-0381-2018
- Recalling firm
- Owen Mumford USA, Inc.
- Product
- Unifine Pentips 5mm x 31G pen needles, packaged in 100 ct pen needles. Product Usage: Sterile, non-toxic, non-pyrogenic, single use pen needles for the subcutaneous injection of drugs from pen injectors.
- Status
- Terminated
- Initiated
- 2017-11-06
- Posted
- not on file
- Terminated
- 2021-02-12
- Reason
- Device may potentially be compromised by water in end user packaging. Exposure to liquid may compromise sterility. This may appear as discoloring or yellowing of the device's protective seal.
- Root cause
- Environmental control
- Action
- Owen Mumford sent an Urgent Medical Device Recall Notification letter dated November 6, 2017 via email. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to check inventory for affected product, contact the Customer Service Department to arrange return of affected product, complete and return the response form and to either provide a customer list for Owen and Mumford to notify their customers or confirmation that they will notify their customers if further distributed.
- Quantity
- 2,660 cartons (266,000 pen needles)
- Distribution
- US Nationwide Distribution in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, MA, MI, MN, MO, MS, NC, NH NJ, NY, OH, OK, PA, TN, NX, UT, VA, VT, WA, WV
- Lot, serial or model codes
- UPC #384703550018: Batch No. AN 3550, Lot No. 1705722.S846; Batch No. AN 3550AC, Lot No. 1705122.S929.
- 510(k) numbers
- ["K152339"]
- PMA numbers
- not on file
- Product code
- FMI
- Firm FEI
- 1000219614
- Firm city
- Marietta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as