Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: REF 200068 Security (TM) Enclosed Carpal Tunnel System Blade ( Single) ABS /Stainless Steel Sterile Device, Single Use Only. Designed to release the transverse carpal ligament.

Recall number
Z-0381-2014
Recalling firm
Biomet, Inc.
Product
REF 200068 Security (TM) Enclosed Carpal Tunnel System Blade ( Single) ABS /Stainless Steel Sterile Device, Single Use Only. Designed to release the transverse carpal ligament.
Status
Terminated
Initiated
2013-10-25
Posted
2013-11-22
Terminated
2015-05-06
Reason
Security Blades were manufactured oversized and potentially won't advance or may be difficult to advance through the mating Security Clip.
Root cause
Nonconforming Material/Component
Action
10/25/2013 Urgent Medical Device Recall Notifications were sent to each Distributor and Risk/Recall Manager at each direct account identifying the recalled device, possible adverse health consequences and instructing consignees to immediately locate and remove devices from circulation, following instructions on the FAX back form and returning it within 3 business days to FAX 574-372-1683 and returning product to Biomet Return Goods, 56 East Bell Drive Warsaw In 46580. Questions should be directed to 574-372-1570 M-F, 8 a.m. to 5 p.m.
Quantity
1130
Distribution
Worldwide Distribution: USA (nationwide) and Internationally to Japan.
Lot, serial or model codes
Catalog Number: 200068 Medical Device Listing Number: B008657 Lot Number Identification: 088610, 090600, 143810, 257060, 289900, 365910, 460490, 558300, 626660, 697320,767260,795090,857730,864640,979680
510(k) numbers
not on file
PMA numbers
not on file
Product code
FZX
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register