Stryker Spine: Stryker MANTIS TORQUE WRENCH REF 48287028 Manufactured by: Stryker pine SAS Z.I MARTICOT-33610-CESTAS-FRANCE +33 (0)5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine, 2 Pearl Ct. Allendale NJ 07401-1677 USA +1-201-780-8000 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached.
- Recall number
- Z-0381-2013
- Recalling firm
- Stryker Spine
- Product
- Stryker MANTIS TORQUE WRENCH REF 48287028 Manufactured by: Stryker pine SAS Z.I MARTICOT-33610-CESTAS-FRANCE +33 (0)5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine, 2 Pearl Ct. Allendale NJ 07401-1677 USA +1-201-780-8000 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached.
- Status
- Terminated
- Initiated
- 2012-08-09
- Posted
- 2012-11-19
- Terminated
- 2015-09-30
- Reason
- Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.
- Root cause
- Nonconforming Material/Component
- Action
- Stryker Spine sent an Urgent Medical Device Removal Notification letter/Product Accountability Forms dated August 9, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. For questions regarding this recall call 201-760-8150.
- Quantity
- 19 units
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Australia, Canada, France, Germany, Netherlands, Japan, Spain, Denmark, Switzerland, and the United Kingdom.
- Lot, serial or model codes
- Class I device, 510(K) exempt Catalog Number: 48287028 Lot Number: 098400
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LXH
- Firm FEI
- 3004024955
- Firm city
- Allendale
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28