Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corp: Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog number 801046.

Recall number
Z-0381-2007
Recalling firm
Terumo Cardiovascular Systems Corp
Product
Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog number 801046.
Status
Terminated
Initiated
2006-11-08
Posted
2007-02-01
Terminated
2007-06-11
Reason
Following installation of software version 1.30 in October 2006, the firm confirmed a complaint that the new software caused a discrepancy in the pump speed calculation and caused the pump to stop during clinical use and a ''connect motor'' message to be displayed.
Root cause
Other
Action
Customers were notified by phone, fax, email or by a Terumo representative of these problems between 11/08/06 and 11/13/06 and instructed to not use the unit until new software had been installed. All version 1.30 software was replaced and version 1.40 installed prior to 12/4/06.
Quantity
35
Distribution
Nationwide.
Lot, serial or model codes
Serial numbers 0119, 0123, 0124, 0128, 0135, 0136, 0139, 0140, 0146, 0147, 0148, 0151, 0152, 0157, 0171, 0196, 0197, 0214, 0232, 0241, 0271, 0305, 0310, 0315, 0316, 0318, 0323, 0345, 0346, 0372, 0375, 0376, 0377, 0378 and 0380.
510(k) numbers
["K022947"]
PMA numbers
not on file
Product code
DTQ
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register