FDA Medical Device Recalls (openFDA)
Terumo Cardiovascular Systems Corp: Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog number 801046.
- Recall number
- Z-0381-2007
- Recalling firm
- Terumo Cardiovascular Systems Corp
- Product
- Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog number 801046.
- Status
- Terminated
- Initiated
- 2006-11-08
- Posted
- 2007-02-01
- Terminated
- 2007-06-11
- Reason
- Following installation of software version 1.30 in October 2006, the firm confirmed a complaint that the new software caused a discrepancy in the pump speed calculation and caused the pump to stop during clinical use and a ''connect motor'' message to be displayed.
- Root cause
- Other
- Action
- Customers were notified by phone, fax, email or by a Terumo representative of these problems between 11/08/06 and 11/13/06 and instructed to not use the unit until new software had been installed. All version 1.30 software was replaced and version 1.40 installed prior to 12/4/06.
- Quantity
- 35
- Distribution
- Nationwide.
- Lot, serial or model codes
- Serial numbers 0119, 0123, 0124, 0128, 0135, 0136, 0139, 0140, 0146, 0147, 0148, 0151, 0152, 0157, 0171, 0196, 0197, 0214, 0232, 0241, 0271, 0305, 0310, 0315, 0316, 0318, 0323, 0345, 0346, 0372, 0375, 0376, 0377, 0378 and 0380.
- 510(k) numbers
- ["K022947"]
- PMA numbers
- not on file
- Product code
- DTQ
- Firm FEI
- 1828100
- Firm city
- Ann Arbor
- Firm state
- MI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28