FDA Medical Device Recalls (openFDA)
RAYSEARCH LABORATORIES AB: RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.
- Recall number
- Z-0380-2018
- Recalling firm
- RAYSEARCH LABORATORIES AB
- Product
- RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.
- Status
- Terminated
- Initiated
- 2017-07-17
- Posted
- not on file
- Terminated
- 2018-08-29
- Reason
- For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used for QA measurements. This could result in an incorrect dosage being provided to a patient.
- Root cause
- Device Design
- Action
- Ray Search sent a Field Safety Notice letter dated July 14, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter included a response form with instructions for workarounds until the software is corrected.
- Quantity
- 117 units
- Distribution
- US Nationwide Distribution - FL, WA, AZ, TN, MD, TX, FL, MI, Washington D.C.
- Lot, serial or model codes
- 6.0.0.24 , 6.1.0.26 , 6.1.1.2 or 6.2.0.7
- 510(k) numbers
- ["K171536"]
- PMA numbers
- not on file
- Product code
- MUJ
- Firm FEI
- 3007774465
- Firm city
- Stockholm
- Firm state
- not on file
- Firm country
- Sweden
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28