Skip to the content

FDA Medical Device Recalls (openFDA)

RAYSEARCH LABORATORIES AB: RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.

Recall number
Z-0380-2018
Recalling firm
RAYSEARCH LABORATORIES AB
Product
RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.
Status
Terminated
Initiated
2017-07-17
Posted
not on file
Terminated
2018-08-29
Reason
For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used for QA measurements. This could result in an incorrect dosage being provided to a patient.
Root cause
Device Design
Action
Ray Search sent a Field Safety Notice letter dated July 14, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. The letter included a response form with instructions for workarounds until the software is corrected.
Quantity
117 units
Distribution
US Nationwide Distribution - FL, WA, AZ, TN, MD, TX, FL, MI, Washington D.C.
Lot, serial or model codes
6.0.0.24 , 6.1.0.26 , 6.1.1.2 or 6.2.0.7
510(k) numbers
["K171536"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
3007774465
Firm city
Stockholm
Firm state
not on file
Firm country
Sweden

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register