Skip to the content

FDA Medical Device Recalls (openFDA)

Microgenics Corporation: Thermo Scientific MAS Omni IMMUNE and Omni IMMUNE PRO; Liquid Assayed Integrated Immunoassay Control. Intended for use as an assayed control for monitoring assay conditions in many clinical laboratory determinations. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance or reagent and instrument. .

Recall number
Z-0379-2013
Recalling firm
Microgenics Corporation
Product
Thermo Scientific MAS Omni IMMUNE and Omni IMMUNE PRO; Liquid Assayed Integrated Immunoassay Control. Intended for use as an assayed control for monitoring assay conditions in many clinical laboratory determinations. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance or reagent and instrument. .
Status
Terminated
Initiated
2012-08-20
Posted
2012-11-19
Terminated
2013-03-04
Reason
In house testing confirmed that CK-MS and Insulin claimed in both Omni IMMUNE and Omni IMMUNE PRO products are displaying reduced stability.
Root cause
Nonconforming Material/Component
Action
Microgenics Corporation sent a Recall Notification letter dated August 20, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed that immediate attention is required and identify the two affected constituents. The letter contains a list of the affected product with their catalog number, lot number and expiration date and request that the CK-MB and Insulin use be discontinued. The remaining components of the product may be used. Evaluate quality controls according to your laboratory procedures. Retain a copy of the letter for your laboratory records. If the product was forwarded to another laboratory, provide them with a copy of the letter. Complete the attached Recall Response form and return within 5 days. Questions may be directed to 1-800-232-3342.
Quantity
2076 total in US , 596 total Foreign
Distribution
Worldwide Distribution--USA (nationwide) and the countries of Germany, Canada and Australia.
Lot, serial or model codes
MAS Onmi IMMUNE: Catalog Number/lot number/ expiration date: OIM-101 - OIM14031 - 31 MAR 2014; OIM-202 - OIM14032 - 31 MAR 2014; OIM-303 - OIM14033 - 31 MAR 2014; OIM-SP - OIM1403S - 31 MAR 2014.  MAS Omni IMMUNE PRO: OPRO-101 - OPRO14031 - 31 MAR 2014; OPRO-201 - OPRO14032 - 31 MAR 2014' OPRO-303 - OPRO14033 - 31 MAR 2014; OPRO-SP - OPRO1403S - 31 MAR 2014.
510(k) numbers
["K110616"]
PMA numbers
not on file
Product code
JJY
Firm FEI
3010939897
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register