FDA Medical Device Recalls (openFDA)
Microgenics Corporation: Thermo Scientific MAS Omni IMMUNE and Omni IMMUNE PRO; Liquid Assayed Integrated Immunoassay Control. Intended for use as an assayed control for monitoring assay conditions in many clinical laboratory determinations. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance or reagent and instrument. .
- Recall number
- Z-0379-2013
- Recalling firm
- Microgenics Corporation
- Product
- Thermo Scientific MAS Omni IMMUNE and Omni IMMUNE PRO; Liquid Assayed Integrated Immunoassay Control. Intended for use as an assayed control for monitoring assay conditions in many clinical laboratory determinations. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance or reagent and instrument. .
- Status
- Terminated
- Initiated
- 2012-08-20
- Posted
- 2012-11-19
- Terminated
- 2013-03-04
- Reason
- In house testing confirmed that CK-MS and Insulin claimed in both Omni IMMUNE and Omni IMMUNE PRO products are displaying reduced stability.
- Root cause
- Nonconforming Material/Component
- Action
- Microgenics Corporation sent a Recall Notification letter dated August 20, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed that immediate attention is required and identify the two affected constituents. The letter contains a list of the affected product with their catalog number, lot number and expiration date and request that the CK-MB and Insulin use be discontinued. The remaining components of the product may be used. Evaluate quality controls according to your laboratory procedures. Retain a copy of the letter for your laboratory records. If the product was forwarded to another laboratory, provide them with a copy of the letter. Complete the attached Recall Response form and return within 5 days. Questions may be directed to 1-800-232-3342.
- Quantity
- 2076 total in US , 596 total Foreign
- Distribution
- Worldwide Distribution--USA (nationwide) and the countries of Germany, Canada and Australia.
- Lot, serial or model codes
- MAS Onmi IMMUNE: Catalog Number/lot number/ expiration date: OIM-101 - OIM14031 - 31 MAR 2014; OIM-202 - OIM14032 - 31 MAR 2014; OIM-303 - OIM14033 - 31 MAR 2014; OIM-SP - OIM1403S - 31 MAR 2014. MAS Omni IMMUNE PRO: OPRO-101 - OPRO14031 - 31 MAR 2014; OPRO-201 - OPRO14032 - 31 MAR 2014' OPRO-303 - OPRO14033 - 31 MAR 2014; OPRO-SP - OPRO1403S - 31 MAR 2014.
- 510(k) numbers
- ["K110616"]
- PMA numbers
- not on file
- Product code
- JJY
- Firm FEI
- 3010939897
- Firm city
- Fremont
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28