Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Corp.: Roche Ammonia/Ethanol/CO2 Control A (Abnormal) for use in conjunction with Roche/Hitachi, COBAS INTEGRA, cobas c111 and cobas 46250 c 501 clinical chemistry analyzers; Cat. Number. 20753009190. Roche Diagnostics Corporation, Indianapolis, IN

Recall number
Z-0379-2008
Recalling firm
Roche Diagnostics Corp.
Product
Roche Ammonia/Ethanol/CO2 Control A (Abnormal) for use in conjunction with Roche/Hitachi, COBAS INTEGRA, cobas c111 and cobas 46250 c 501 clinical chemistry analyzers; Cat. Number. 20753009190. Roche Diagnostics Corporation, Indianapolis, IN
Status
Terminated
Initiated
2007-11-14
Posted
2007-12-29
Terminated
2009-09-22
Reason
High Quality Control Recovery: Undetected high quality control recovery in bicarbonate assay.
Root cause
Process design
Action
Consignees were notified of this recall by an Urgent Medical Device Recall letter sent on 11/14/07. The letter instructed the users to immediately discontinue use of Ammonia/ Ethanol/CO2 Control N and Ammonia/Ethanol/CO2 Control A (with listed lots) for quality control assessment of the bicarbonate assay. They were informed that controls may still be used to assess ammonia and ethanol, but alternative controls must be used for assessment of the bicarbonate assay. Consignees are instructed to destroy products at their facility. Consignees were notified of the availability of replacement product via letter dated .
Quantity
1,296
Distribution
Nationwide Distibution
Lot, serial or model codes
Lots No: 179021-01 and 179021-02; exp. 3/31/2008.
510(k) numbers
["K031880"]
PMA numbers
not on file
Product code
JJY
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register