Skip to the content

FDA Medical Device Recalls (openFDA)

AtriCure, Inc.: AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip Product is used for the occlusion of the heart's left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures.

Recall number
Z-0377-2017
Recalling firm
AtriCure, Inc.
Product
AtriClip LAA Exclusion System with preloaded Gillinov-Cosgrove Clip Product is used for the occlusion of the heart's left atrial appendage, under direct visualization, in conjunction with other open cardiac surgical procedures.
Status
Terminated
Initiated
2016-09-22
Posted
not on file
Terminated
2018-02-15
Reason
There is a potential for a component failure of the device which could necessitate medical intervention.
Root cause
Device Design
Action
AtriCure sent an Urgent Advisory Notice dated September 21, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately examine their inventory and determine if they have any affected product. If so customers should ensure that all users of the affected devices are aware of the Advisory Notice. If customers further distributed the product, they should identify their customers and notify them at once of the Advisory Notice. Any units from the affected lots should be returned to AtriCure, Inc. Customers should contact AtriCure customer service at 1-866-349-2342, and select option 6 to return any affected product(s). Customers were asked to complete and return the enclosed Device Notification Acknowledgement Form as soon as possible. Customers with questions should call 513-755-4563.
Quantity
298
Distribution
Nationwide Distribution to the follwoing states: AL, AR, AZ, CA, CT, FL, GA, IL, KS, MD, MI, MN, MO, MS, NC, NM, NY, OH, OK, PA, SC, TN, TX, VA, VT & WI. Product was also shipped to the following countries: Poland and Switzerland
Lot, serial or model codes
Model Numbers: PRO235, PRO240, PRO245 and PRO250  Lot Numbers: 65622 to 66773
510(k) numbers
["K160454"]
PMA numbers
not on file
Product code
FZP
Firm FEI
3011706110
Firm city
Mason
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register