Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Electronics North America Corporation: Philips Healthcare System X-ray Tomography Computed :

Recall number
Z-0376-2017
Recalling firm
Philips Electronics North America Corporation
Product
Philips Healthcare System X-ray Tomography Computed :
Status
Terminated
Initiated
2016-03-15
Posted
not on file
Terminated
2020-09-09
Reason
It was discovered the patient support tabletop may fail to move in the horizontal direction when it is commanded to do so during CT scans due to a potential mechanical linkage failure.
Root cause
Radiation Control for Health and Safety Act
Action
Philips planned action: 1. Philips issued a customer notification letter on October 13, 2016. The letter includes a statement that the corrections will be made free of charge. 2. Philips will initiate a field corrective action to address the issue. 3. Philips will contact the customer and schedule an inspection to determine if the unit requires repair. 4. If repair is required, the couch horizontal linkage assembly will be replaced. For further questions please call (978) 659-3000.
Quantity
1363
Distribution
USA(nationwide) Distribution to the states of : AZ, NY, GA, UT, TN, TX, AR, NV, Il, FL, MS, LA, MD, MO, MA, OK, KY, WY, SC, ND, NE, DE, MI, WA, CT, RI, KS, OR, IN, PA, CA, NJ, NC, MN, AL and IA.
Lot, serial or model codes
Ingenuity CT, Ingenuity Core, Ingenuity Core 128.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JAK
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register