Stryker Spine: Stryker 1) XIA Torque Wrench 2) XIA 3 Torque Wrench 3) XIA Elegance Short Torque Wrench Manufactured by: Stryker Spine SAS Z.I. Marticot -33610 -CESTAS -FRANCE +33 (0) 5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine 2 Pearl Court Allendale, NJ 07401-1677 USA +1-201-760-8000 XIA 3 Spinal System- Intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, non cervical pedicle and on-pedicled fixation system, the XIA 3 Spinal system is intended to provide additional support during fusion autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities; degenerate disc disease (DDD) ; spondylolisthesis; trauma spinal stenosis; curvatures; tumor; pseudoarthrosis and failed previous fusion. XIA - Intended for anterior/anterolateral and posterior, non cervical pedicle and non-pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease; spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudoarthrosis and failed previous fusion.

Recall number
Z-0375-2014
Recalling firm
Stryker Spine
Product
Stryker 1) XIA Torque Wrench 2) XIA 3 Torque Wrench 3) XIA Elegance Short Torque Wrench Manufactured by: Stryker Spine SAS Z.I. Marticot -33610 -CESTAS -FRANCE +33 (0) 5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine 2 Pearl Court Allendale, NJ 07401-1677 USA +1-201-760-8000 XIA 3 Spinal System- Intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, non cervical pedicle and on-pedicled fixation system, the XIA 3 Spinal system is intended to provide additional support during fusion autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities; degenerate disc disease (DDD) ; spondylolisthesis; trauma spinal stenosis; curvatures; tumor; pseudoarthrosis and failed previous fusion. XIA - Intended for anterior/anterolateral and posterior, non cervical pedicle and non-pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease; spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudoarthrosis and failed previous fusion.
Status
Terminated
Initiated
2013-09-24
Posted
2013-11-21
Terminated
2016-10-13
Reason
Multiple complaints have reported that the hex tip of the torque wrench fracturing during final tightening.
Root cause
Process design
Action
Stryker sent an Urgent Medical Device Removal Notification letter dated September 25, 2013, with product accountability forms to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory and hospital locations to identify any affected product, reconcile any Torque Wrenches with Catalog Number with Lot Numbers referenced in the letter. Customers were asked to fax a copy of the Customer Response Form to Regulatory Compliance at 855-632-9049 or you email it to Spine-RegulatoryActions@Stryker.com. Customers were asked to return the affected product using the enclosed mailing label to: RECALL RA 2013-105 Stryker Spine 59 Route 17 South, Allendale, New Jersey 07401 branches/agencies, the field and surgeons via Fed Ex on starting on 9/24/2013. For questions regarding this recall call 201-760-8298. Stryker sent another Urgent Medical Device Removal Notification Letter dated 1/29/2014 which included additional lots and one product.
Quantity
220 US units; 431 OUS units
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia, Canada, Chile, France, Canada, Germany, Hong Kong, Ireland, Italy, Japan, Korea, Netherlands, New Zealand, Spain, Iraq, Egypt, and Turkey.
Lot, serial or model codes
Medical Device Listing Number E207315  1) XIA Torque Wrench Catalog # 03807028 Lot # 127510, 092793, 097089, 11E038, 11E039, 11E040, 11E041, 123746, 127179, &12D012  2) XIA 3 Torque Wrench Catalog # 48237028 Lot # 11E042, 11E044, 11E047, 098525, 118823, 11E043, 11E045, 11E046, 11E048, 127647, R127647, 12A646, & R12A646  3) XIA Elegance Short Torque Wrench Catalog # 482397028 Lot # 11A957, 11E035, 11E036, R11E036, & 121098  4) Mantis Redux Torque Wrench Reference Number 48287028 Lot # 098401, 125708, 125709, 127051 & 12A641
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
3004024955
Firm city
Allendale
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register