Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Point of Care Inc.: Abbott Point of Care, i-STAT G Cartridge; (Glucose/Glu); List No: 06F09-01; Part No: 120100; Abbott Point of Care, Inc., East Windsor, NJ 08520, USA; Product of Canada.

Recall number
Z-0375-2008
Recalling firm
Abbott Point of Care Inc.
Product
Abbott Point of Care, i-STAT G Cartridge; (Glucose/Glu); List No: 06F09-01; Part No: 120100; Abbott Point of Care, Inc., East Windsor, NJ 08520, USA; Product of Canada.
Status
Terminated
Initiated
2007-10-18
Posted
2007-12-11
Terminated
2008-07-18
Reason
Difficult to Close -- Certain lots of i-STAT cartridges are difficult to close, will not close or do not stay closed.
Root cause
Packaging change control
Action
Abbott Point of Care sent and Urgent Notice to all i-STAT Customers (US and Canadian consignees) notifying them that certain lots of i-STAT cartridges have snap closures that are difficult to close, will not close or do not stay closed. The customers were asked to review the lot numbers on the enclosure and, if the facility has any cartridges from the lot numbers listed, they should remove from service immediately.
Quantity
not on file
Distribution
Worldwide Distribution- including USA and country of Canada.
Lot, serial or model codes
Catalog Number: 06F09-01; 06F09-02 Lot Numbers: A07102A; A07124; A07138; D07117; D07119; D07120; and D07121.
510(k) numbers
["K001154","K894914","K912387"]
PMA numbers
not on file
Product code
JGS
Firm FEI
2245578
Firm city
East Windsor
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register