Skip to the content

FDA Medical Device Recalls (openFDA)

Cincinnati Sub-Zero Products: Warm Air Model 135 Warming Units.

Recall number
Z-0374-05
Recalling firm
Cincinnati Sub-Zero Products
Product
Warm Air Model 135 Warming Units.
Status
Terminated
Initiated
2004-11-05
Posted
2004-12-30
Terminated
2008-10-29
Reason
Burning smell, burned microprocessor board. The firm is modifying the connectors and wiring gauge configuration of the device.
Root cause
Other
Action
The firm notified consignees by certified letter on November 5, 2004.
Quantity
649 units.
Distribution
The product was distributed throughout the United States and to foreign countries including Saudi Arabia, Taiwan, Japan, and Canada.
Lot, serial or model codes
All serial numbers.
510(k) numbers
["K942599"]
PMA numbers
not on file
Product code
BTE
Firm FEI
1516825
Firm city
Cincinnati
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register