Medtronic Neuromodulation: Medtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used with Neurostimulator Inceptiv INS Model 977119 and Percept RC INS Model B35300 as part of the Deep Brain Stimulation (DBS) system. WR9220A is used with InterStim Micro, Model 97810, as part of Pelvic Health system

Recall number
Z-0372-2025
Recalling firm
Medtronic Neuromodulation
Product
Medtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used with Neurostimulator Inceptiv INS Model 977119 and Percept RC INS Model B35300 as part of the Deep Brain Stimulation (DBS) system. WR9220A is used with InterStim Micro, Model 97810, as part of Pelvic Health system
Status
Open, Classified
Initiated
2024-09-18
Posted
2024-11-13
Terminated
not on file
Reason
The Wireless Rechargers in the kits may be unable to enter a recharging session upon first use.
Root cause
Under Investigation by firm
Action
The recalling firm issued letters dated September 2024 on 9/18/2024 via UPS 2-Day Air. The letter explained the issue, product scope, and the customer actions. The actions included: (1) Identify and quarantine all unused affected WR Kits. Attachment A of the letter provides guidance on identifying potentially affected devices; (2) Return all unused affected WR Kits using the instructions in the enclosed customer confirmation form and contact Medtronic Customer Service to obtain a replacement WR Kit; (3) Direct patients who cannot establish a recharge session using their WR to promptly contact Medtronic Customer Service to obtain a replacement WR Kit; (4) Complete the enclosed Customer Confirmation Form; (5) Share this notice with all those who need to be aware of this issue within or outside your organization or to any organization where the potentially affected product has been transferred or distributed and maintain a copy of this notice in your records. Medtronic initiated communication of the potential issue via physical mailing to one (1) customer beginning on 08-July-2025 in the U.S.
Quantity
118 wireless rechargers
Distribution
Worldwide - US Nationwide distribution in the states of AL, AZ, CA, FL, GA, ID, MD, MI, MN, MO, NC, NJ, OH, PA, SC, TX, VA, and WA. The countries of Austria, Belgium, Denmark, Finland, Germany, Ireland, Italy, Japan, Netherlands, Norway, Portugal, Spain, Sweden, Switzerland, and United Kingdom.
Lot, serial or model codes
RS6230 - GTIN 00763000564353, Serial numbers: NRA010637N; NRA011389N; NRA013260N; NRA013664N; NRA013924N; NRA014318N; NRA014347N; NRA015823N; NRA015861N; NRA015864N; NRA016334N; NRA016413N; NRA016430N; NRA016432N; NRA016778N; RA017002N; NRA017048N; NRA017227N; NRA017232N; NRA017685N; NRA018697N; NRA018731N; NRA018906N; NRA018936N; NRA018992N; NRA019019N; NRA019026N; NRA019963N; NRA019967N; NRA019968N; NRA019970N; NRA019983N; NRA010657N; NRA012367N; NRA012898N; NRA012915N; NRA013490N; NRA013643N; NRA013877N; NRA013878N; NRA013880N; NRA014248N; NRA014741N; NRA015127N; NRA015142N, NRA010659N; NRA010661N; NRA010663N NRA016424N; NRA016425N; NRA016426N; NRA016431N; RA016433N; NRA021324N. RS7230 - GTIN 00763000540289, Serial numbers: NRA010640N; NRA010650N; NRA010656N: NRA011591N; NRA012896N; NRA012910N; NRA012925N; NRA013884N; NRA010633N; NRA010635N; NRA010641N; NRA010642N; NRA010646N; NRA010652N; NRA010653N; NRA010654N; NRA010655N; NRA010658N; NRA010660N; NRA010664N; NRA010666N; NRA010667N; NRA010668N; NRA010669N; NRA010881N; NRA011216N; NRA011293N; NRA011393N; NRA011713N; NRA011853N; NRA012184N; NRA012897N; NRA012909N; NRA012911N; NRA012912N; NRA012913N; NRA012914N; NRA013502N; NRA013602N; NRA013603N; NRA013634N; NRA013635N; NRA013638N; NRA013639N; NRA013879N; NRA013883N; NRA013899N; NRA013952N; NRA014259N; NRA014944N; NRA015858N; NRA015904N; NRA011189N; NRA011238N; NRA012130N; NRA012185N; NRA012366N; NRA013543N; NRA014132N; NRA014222N; GTIN 0763000827533, Serial numbers: NRA010558N; NRA010627N; NRA015360N; NRA013633N.
510(k) numbers
not on file
PMA numbers
["P840001","P960009","P970004","P080025"]
Product code
LGW
Firm FEI
2182207
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register