FDA Medical Device Recalls (openFDA)
Hospira Inc.: *** 1) SYMBIQ One-Channel Infuser; *** 2) SYMBIQ Two-Channel Infuser; *** E09 batteries can be identified by the "E09" that is printed beneath the bar code and at the end of the serial number along the bottom of the label found on each battery; *** E10 batteries can be identified by the "E10" that is printed beneath the bar code and at the end of the serial number along the bottom of the label found on each battery. Symbiq is a general-purpose infuser designed to deliver fluids, solutions, medications, agents, nutritionals, electrolytes, blood and blood products for parenteral, enteral, intra-arterial, subcutaneous, or irrigation routes of administration.
- Recall number
- Z-0372-2014
- Recalling firm
- Hospira Inc.
- Product
- *** 1) SYMBIQ One-Channel Infuser; *** 2) SYMBIQ Two-Channel Infuser; *** E09 batteries can be identified by the "E09" that is printed beneath the bar code and at the end of the serial number along the bottom of the label found on each battery; *** E10 batteries can be identified by the "E10" that is printed beneath the bar code and at the end of the serial number along the bottom of the label found on each battery. Symbiq is a general-purpose infuser designed to deliver fluids, solutions, medications, agents, nutritionals, electrolytes, blood and blood products for parenteral, enteral, intra-arterial, subcutaneous, or irrigation routes of administration.
- Status
- Terminated
- Initiated
- 2013-08-28
- Posted
- 2013-11-20
- Terminated
- 2016-11-07
- Reason
- Hospira shipped the incorrect replaceable battery pack (Part No: HSP3566-E10) instead of the correct E09 battery pack (Part No. HSP3566-E09) for the Symbiq Single and Dual Channel Infusers.
- Root cause
- Employee error
- Action
- The firm, Hospira, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated August 28, 2013 to its customers. Hospira contracted with Stericycle, Inc., in Indianapolis, IN for customer notification, response, tracking, returns, effectiveness checks and destruction of returned product starting on August 28, 2013. The letter described the product, problem and actions to be taken. Customers were instructed to: 1) complete and return the attached reply form via fax to 1-866-470-3291 or e-mail to Hospira2921@stericycle.com, even if you do not currently have the affected batteries; 2) contact Stericycle at 1-877-779-7810 (M-F, 8am - 5pm ET) if they require additional response forms; and, 3) return any affected batteries upon receipt of the new (correct) replacement batteries to Stericycle. Customers are also instructed to forward the recall information to any accounts that they may have shipped the affected batteries to and have them contact Stericycle to recieve a reply form and return label. For further inquiries, please contact Hospira Advanced Knowledge Center at 1-800-241-4002, option 6, availabel 24 hours a day/7 days per week.
- Quantity
- 59 batteries
- Distribution
- US Distribution in states of: : KY, MN, ND, NJ, NY, PA, and SD.
- Lot, serial or model codes
- *** 1) SYMBIQ One-Channel Infuser, List Number: 16026-04-01/02/03/04/25/26/43/44/45/46/51/52/53/54/55/56/57/58/65/66/69/70/77/78/79/80/81/82/83/84/87/88/89/90/93/94; *** 2) SYMBIQ Two- Channel Infuser, List Number: 16027-04-01/02/03/04/25/26/43/44/45/46/51/52/53/54/55/56/57/58/65/66/69/70/77/78/79/80/81/82/83/84/87/88/89/90/ 93/94; *** X) Battery Pack E10 Part Number: HSP3566-E10; *** X) Battery Pack E09 Part Number: HSP3566-E09
- 510(k) numbers
- ["K110901"]
- PMA numbers
- not on file
- Product code
- FRN
- Firm FEI
- 3013319212
- Firm city
- Lake Forest
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28