Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: GAMBRO Capillary Dialyzer, Polyflux(R) Revaclear MAX, STERILE, Product code 110634 Product Usage: The Revaclear Max dialyzer is a single use device used for treatment of chronic and acute renal failure by hemodialysis.

Recall number
Z-0370-2017
Recalling firm
Baxter Healthcare Corp.
Product
GAMBRO Capillary Dialyzer, Polyflux(R) Revaclear MAX, STERILE, Product code 110634 Product Usage: The Revaclear Max dialyzer is a single use device used for treatment of chronic and acute renal failure by hemodialysis.
Status
Terminated
Initiated
2016-10-12
Posted
2016-11-08
Terminated
2017-08-19
Reason
Potential presence of particulate matter on the blood side of the dialyzer
Root cause
Process control
Action
Baxter sent an Urgent Product Recall letter dated October 12, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to: I. Locate and remove all affected product lots from their facility. 2. Contact Baxter Healthcare Center for Service to arrange for return and credit. 3. Complete the enclosed Baxter Customer Reply Form and return it to Baxter by either fax or scanned email. For questions contact Baxter Product Surveillance at 800-437-5176.
Quantity
22,656 units
Distribution
Worldwide Distribution - US Nationwide and the countries of Canada and Mexico.
Lot, serial or model codes
C416201601
510(k) numbers
["K060195"]
PMA numbers
not on file
Product code
KDL
Firm FEI
1417572
Firm city
Deerfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register