GE Healthcare, LLC: GE LightSpeed VCT Scanner System, models 5212920-300, 5212920-310, 5311595-10. Composed of a gantry, patient table, operator console, computer, and PDU and includes image acquisition hardware, image acquisition and reconstruction software, associated accessories and connections/interfaces to accessories. GE Healthcare, 3000 North Grandview, Waukesha, WI 53188 USA. Is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions.

Recall number
Z-0368-2011
Recalling firm
GE Healthcare, LLC
Product
GE LightSpeed VCT Scanner System, models 5212920-300, 5212920-310, 5311595-10. Composed of a gantry, patient table, operator console, computer, and PDU and includes image acquisition hardware, image acquisition and reconstruction software, associated accessories and connections/interfaces to accessories. GE Healthcare, 3000 North Grandview, Waukesha, WI 53188 USA. Is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions.
Status
Terminated
Initiated
2010-09-03
Posted
2010-11-15
Terminated
2012-05-15
Reason
GE has become aware of a potential set of circumstances that could cause x-ray continuation during an unexpected table stop on certain LightSpeed VCT and BrightSpeed systems.
Root cause
Software design
Action
An Urgent Medical Device Correction letter dated September 2, 2010 was sent to consignees beginning September 3, 2010. This letter addressed the LightSpeed VCT systems. Another Urgent Medical Device Correction letter, dated September 24, 2010 was sent September 24, 2010 for the BrightSpeed Systems. The letters described the issue and provided safety instructions for the health care provider to be aware if the table should stop unexpectedly and the x-ray remains on to manually stop the scan until the devices software is upgraded. If you have any questions, please contact the US (800) 437-1171, Japan 0120-055-919, Korea 1544-6119, Australia/New Zealand 800 659 465, China 800-810-8188; for other countries, please contact your local GE Healthcare Services Representative.
Quantity
68
Distribution
Worldwide distribution: USA including states of: MI, NM, OH, TX, and Puerto Rico and countries including: Algeria, Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guatemala, Honduras, Hungary, India, Indonesia, Ireland, Islamic Republic of Iran, Israel, Italy, Jamaica, Japan, Kenya, Lebanon, Malaysia, Mauritania, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Pakistan, Paraguay, Philippines, Poland, Portugal, Republic of Korea, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Togo, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Viet Nam.
Lot, serial or model codes
Model 5212920-300, serial numbers: 00000021775YC2. 00000408285CN5. 00000409136CN9. 00000412029CN1. 00000412391CN5. 00000414803CN7. 00000415599CN0. 00000416813CN4. 00000417885CN1.   Model 5212920-310, serial numbers: 00000416740CN9. 00000417311CN8. 00000417519CN6. 00000417884CN4. 00000417886CN9. 00000417916CN4. 00000417917CN2. 00000417918CN0. 00000417919CN8. 00000417925CN5. 00000417926CN3. 00000417989CN1. 00000417990CN9. 00000418000CN6. 00000418055CN0. 00000418056CN8. 00000418057CN6. 00000418058CN4. 00000418059CN2. 00000418060CN0. 00000418062CN6. 00000418063CN4. 00000418064CN2. 00000418065CN9. 00000418126CN9. 00000418141CN8. 00000418142CN6. 00000418179CN8. 00000418218CN4. 00000418219CN2. 00000418220CN0. 00000418221CN8. 00000418222CN6. 00000418246CN5. 00000418260CN6. 00000418261CN4. 00000418266CN3. 00000418272CN1. 00000418293CN7. 00000418296CN0. 00000418337CN2. 00000418355CN4. 00000418365CN3. 00000418400CN8. 00000418404CN0. 00000418467CN7. 00000418541CN9. 00000418542CN7. 00000418581CN5. 00000418586CN4. 00000418604CN5. 00000418706CN8. 00000418717CN5. 00000418789CN4. 00000418790CN2. 00000418793CN6.   Model 5311595-10, serial numbers: 00000031546CO5. 00000031835CO2. 00000031838CO6.
510(k) numbers
["K052855","K082761","K082816"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register