Skip to the content

FDA Medical Device Recalls (openFDA)

Merz North America, Inc.: PROLARYN Injectable Implant 1.0 cc Gel, Model No. 8602M0K5. A restorable implant material to aid in surgical reconstructions as a space occupying material in laryngeal surgical procedures for vocal fold medialization and augmentation.

Recall number
Z-0367-2016
Recalling firm
Merz North America, Inc.
Product
PROLARYN Injectable Implant 1.0 cc Gel, Model No. 8602M0K5. A restorable implant material to aid in surgical reconstructions as a space occupying material in laryngeal surgical procedures for vocal fold medialization and augmentation.
Status
Terminated
Initiated
2015-10-26
Posted
2015-12-03
Terminated
2017-05-08
Reason
The expiration date encoded in the UDI barcode on the label on the product carton and the peelable patient identification label is incorrect.
Root cause
Other
Action
Letters were issued 11/4/2015 to consignees. The letter informs the consignees that they can continue to safely use the affected products by referring to the USE BEFORE date. They letter does not require for any product to be returned. Also, there is no sub-recall required as the product was not distributed to distributors or wholesalers.
Quantity
7,715 units
Distribution
Distributed US (nationwide) including the Virgin Islands and South Africa.
Lot, serial or model codes
Lot Number 100079214, 100077099, 100077251, 100077252, 100077253, 100078692, 100079598, 100080204, 100081409, 100081577, 100082304, 100082438, 100083371, 100083875, 100083932
510(k) numbers
["K060815"]
PMA numbers
not on file
Product code
MIX
Firm FEI
1012187
Firm city
Raleigh
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register