Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Inc.: Zimmer ZMR Femoral Spout & Cone Bodies and Trabecular Metal Hip Prosthesis Femoral Stems: FEMORAL BODY REVISION NITRIDED POROUS 12/14 NECK TAPER EXTENDED 46 MM NECK OFFSET SPOUT BODY SIZE E 55 MM BUILD-UP TIVANIUM TI-6AL-4V ALLOY STERILE. Item number: 00-9991-021-55

Recall number
Z-0367-2012
Recalling firm
Zimmer Inc.
Product
Zimmer ZMR Femoral Spout & Cone Bodies and Trabecular Metal Hip Prosthesis Femoral Stems: FEMORAL BODY REVISION NITRIDED POROUS 12/14 NECK TAPER EXTENDED 46 MM NECK OFFSET SPOUT BODY SIZE E 55 MM BUILD-UP TIVANIUM TI-6AL-4V ALLOY STERILE. Item number: 00-9991-021-55
Status
Terminated
Initiated
2011-11-16
Posted
not on file
Terminated
2013-08-28
Reason
Lack of assurance of proper metal fatigue strength due to metal grain structure anomaly. In 2006 a recall was initiated as a result of the grain anomaly; however, these 4 lots had been split off during normal production and had not been included in the 2006 recall.
Root cause
Nonconforming Material/Component
Action
On 11/16/11, Zimmer sent letters titled "URGENT MEDICAL DEVICE RECALL" to distributors via electronic mail and to risk managers and surgeons via FedEx indicating that the product must be immediately removed from use and quarantined. " Stop using the affected lots and place affected product in quarantine. " Review available product on the shelf " Identify those part/lot combinations listed in the recall communication " Immediately remove any implants found that match part/lot combinations listed in the recall communication and quarantine " Return affected lots and signature of acknowledgement to Zimmer
Quantity
1 unit (International Only Switzerland)
Distribution
Worldwide distribution: USA (NY), AUSTRALIA, SWITZERLAND, and DENMARK
Lot, serial or model codes
Item number: 00-9991-021-55 Lot number: 00104623
510(k) numbers
["K994286"]
PMA numbers
not on file
Product code
LPH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register