Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog Number: (1) L2KPP2, (2) L2KPP6;

Recall number
Z-0366-2026
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog Number: (1) L2KPP2, (2) L2KPP6;
Status
Open, Classified
Initiated
2025-09-22
Posted
2025-10-27
Terminated
not on file
Reason
The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on the IMMULITE 2000/IMMULITE 2000 XPi systems. The bias is observed in both serum and plasma samples.
Root cause
Under Investigation by firm
Action
On September 22, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Customer Actions: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Discontinue use of and discard the kit lot listed in the table above (Products Section). Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. For continued Intact PTH testing in your laboratory during this time, please contact your Siemens representative to discuss alternative Siemens Healthineers solutions. As an alternative, customers who have an ADVIA Centaur system or an Atellica IM/CI may utilize the corresponding iPTH assay on those systems. These platforms are not impacted. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. A follow-up communication will be provided when assay performance has been restored. ***Update 2/18/2026*** On February 10, 2026 Follow-up Urgent Medical Device Correction letters were sent to additional customers to notify of the addition of new lots for the recall. Customer actions remained the same as the original letter.
Quantity
3318 units (169 US, 3149 OUS)
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, DC, FL, ID, IN, LA, MD, NC, NJ, NY, PA, SC and the countries of Argentina, Australia, Bolivia, Bosnia Herzeg, Brazil, Bulgaria, Canada, Chile, Colombia, Cyprus, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, Hungary, India, Iran, Israel, Italy, Kosovo, Kuwait, Latvia, Lebanon, Macedonia, Mexico, Nepal, Netherlands, P.R. China, Pakistan, Palestinian Ter, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Serbia, Slovakia, Slovenia, South Africa, Spain, Taiwan, Tunisia, Turkey, Turkmenistan, U.A.E., Ukraine, Uruguay, Uzbekistan.
Lot, serial or model codes
1. Siemens Material Number (SMN): 10381441; Catalog Number: L2KPP2; UDI-DI: 00630414961910; Lot Number: 385, ***Expanded 2/18/2026*** 387, 388, 389;  2. Siemens Material Number (SMN): 10381442; Catalog Number: L2KPP6; UDI-DI: 00630414961927; Lot Number: 385, ***Expanded 2/18/2026*** 387, 388, 389;
510(k) numbers
["K013566"]
PMA numbers
not on file
Product code
CEW
Firm FEI
1219913
Firm city
East Walpole
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register