FDA Medical Device Recalls (openFDA)
Smith & Nephew, Inc.: smith&nephew EVOS CONDYLAR MEDIAL DISTAL FEMUR PLATE, R, 5H, 115MM, REF 72573202
- Recall number
- Z-0366-2022
- Recalling firm
- Smith & Nephew, Inc.
- Product
- smith&nephew EVOS CONDYLAR MEDIAL DISTAL FEMUR PLATE, R, 5H, 115MM, REF 72573202
- Status
- Terminated
- Initiated
- 2021-11-01
- Posted
- not on file
- Terminated
- 2026-06-15
- Reason
- Testing vendor notified Smith & Nephew that previous MRI scanning conditions provided in Exponents reports were erroneous. Erroneous data from the reports was incorporated in the released IFU 81110294 EVOS Large Fragment Plating System.
- Root cause
- Under Investigation by firm
- Action
- The firm initiated the correction by email and letter on 1 NOV 2021. The notice explained the problem and requested that the consignees check their inventory and complete the Inventory Correction Certificate form. Distributors were directed to notify their customers of the field action and ensure that these actions are carried out.
- Quantity
- 41 units
- Distribution
- US Nationwide distribution in the states of CO, FL, GA, MA, MN, MO, NC, NJ, NV, NY, OH, OK, PA, TN, TX, WA.
- Lot, serial or model codes
- Batch Number: 20HM10258, 20HM10259, 20HM10261, 20HM22791, UDI
- 510(k) numbers
- ["K201918"]
- PMA numbers
- not on file
- Product code
- HRS
- Firm FEI
- 1020279
- Firm city
- Memphis
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28