Visx Inc: Advanced Medical Optics (AMO), VISX WaveScan WaveFront System (software). Part Number 0070-1478. Advanced Medical Optics, Santa Clara, CA 95051.

Recall number
Z-0365-2008
Recalling firm
Visx Inc
Product
Advanced Medical Optics (AMO), VISX WaveScan WaveFront System (software). Part Number 0070-1478. Advanced Medical Optics, Santa Clara, CA 95051.
Status
Terminated
Initiated
2007-05-30
Posted
2008-01-23
Terminated
2008-04-17
Reason
Erroneous Treatment Calculations - Two software caused errors; in combination, in the WaveScan System Software, will result in an erroneous treatment calculation (overcorrection) in patients. (1) Installation of software for a certain brand of computer (Dedicated Computing LLC) can cause a software registry setting for an algorithm that identifies image reflection to be erroneously set in the "off
Root cause
Software design
Action
AMO sent an Action Required notification on May 30, 2007, to AMO Certified Physicians. The letter included a CD software update (0070-1721) to be installed by users that provided correct registry parameter settings. Upon completion of installation, a confirmation form will print out. Customers are requested to fax back confirmation to AMO (408-773-7055).
Quantity
196 systems are affected in the U.S. and 133 outside of the U.S.
Distribution
Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO,MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK,OR,PA, SC, TN, TX, UT, VA, WA, WI, and WV and countries of Canada and Mexico.
Lot, serial or model codes
System Serial Numbers affected include: 75447, 75448, 76086, 75389, 70170, 76092, 76004, 76020, 76070, 76128, 75391, 75413, 75416, 75417, 75423, 75441, 75442, 76043, 76046, 76048, 76082, 76091, 76121, 76144, 76147, 75419, 76122, 75257, 75385, 75386, 76003, 76022, 76111, 76018, 75390, 76050, 76098, 75420, 76071, 76075, 76114, 75411, 75425, 76123, 76124, 76136, 75384, 75421, 76002, 76023, 75449, 76084, 76119, 76085, 76033, 76036, 76083, 76076, 76052, 76089, 76067, 75444, 75450, 76047, 76099, 76146, 76026, 75424, 76095, 75410, 76106, 75392, 75409, 75428, 75452, 76025, 76032, 76045, 76053, 76097, 70213, 70256, 70316, 70776, 75358, 75418, 75430, 75431, 75432, 76042, 76044, 76049, 76051, 76077, 76078, 76108, 76109, 76116, 76126, 76141, 76035, 76087, 76112, 75398, 76115, 76013, 76120, 76057, 75387, 76021, 76125, 76127, 76130, 76135, 75422, 76069, 75415, 76110, 76137, 76161, 76163, 70419, 70869, 75403, 76096, 76142, 70304, 70160, 70250, 70666, 70753, 70890, 75028, 75402, 75405, 75426, 75429, 75436, 76007, 76027, 76029, 76031, 76037, 76065, 76068, 76073, 78074, 76079, 76103, 76117, 76139, 76140, 76143, 76150, 76151, 76153, 76155, 76156, 76157, 76159, 76160,76162, 70272, 70280, 70297, 70706, 70872, 75297, 75359, 75404, 76059, 76061, 76063, 76081, 76149, 75397, 75399, 70541, 70774, 70904, 75412, 75437, 76017, 70043, 70394, 75401, 75439, 76066, 76080, 76088, 76148, 76058, 76090, 70271, 70353, 70398, 70409, 70734, 76006, 76009, 76028, 76154, 76158, 76145, 70855, 76038, 76062, 75433, 70864, 75394, 75434, 76138, 71704, 70217, 70485, 70491, 70690, 75281, 76104, 75406, 75414, 75427, 76015, 70286, 70440, 70476, 70765, 70882, 75400, 75446, 76005, 75388, 76024, 76039, 76129, 70588, 70603, 75445, 76040, 70597, 70175, 70178, 76094, 70759, 70336, 70637, 75230, 76093, 76105, 75121, 76019, 76056, 76060, 70333, 70341, 70365, 70379, 70428, 75113, 76001, 76054, 70691, 75109, 76107, 70299, 75234, 70554, 70653, 75236, 76011, 76152, 70474, 70495, 70542, 75440, 76101, 76118, 76113, 75435, 76012, 70031, 70032, 70036, 70092, 70227, 70580, 70608, 70721, 75172, 75395, 75408, 75443, 75451, 76016, 76030, 76034, 76072, 76100, 76064, 70080, 70221, 70283, 75328, 75396, 75438, 75407, 70159, 70241, 70350, 76008, 76010, 76014, 70650, 70732, 76055, 76102, 70457, and 76041
510(k) numbers
not on file
PMA numbers
["P930016/S26","P930016S016"]
Product code
LZS
Firm FEI
1221842
Firm city
Santa Clara
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register