Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: Juggerknotless Drill Bit, REF 110003173, Sterile, Single Use Product Usage: The Juggerknotless Soft Anchors are indicated for soft tissue reattachment procedures in the shoulder, foot/ankle, hand/wrist, and hip.

Recall number
Z-0364-2014
Recalling firm
Biomet, Inc.
Product
Juggerknotless Drill Bit, REF 110003173, Sterile, Single Use Product Usage: The Juggerknotless Soft Anchors are indicated for soft tissue reattachment procedures in the shoulder, foot/ankle, hand/wrist, and hip.
Status
Terminated
Initiated
2013-10-10
Posted
2013-11-19
Terminated
2014-07-16
Reason
Investigation identified that the design of the drill bit may make the suture anchor difficult to insert into the bone. The implant may skip across the hole or miss the hole during insertion.
Root cause
Device Design
Action
Biomet sent an Urgent Medical Device Recall Notice letter dated October 10, 2013 to all consignees. The letter identified the affected product, reason for recall, and actions to be taken. The letter stated customers will be contacted on an individual basis in an attempt to reconcile the affected product. Customers were directed to locate and remove implicated devices and fax a copy of the response form to 574-372-1683 prior to returning product. For questions call 574-372-1570.
Quantity
25
Distribution
USA Nationwide Distribution in the states of VA, CA, Ky, and IN
Lot, serial or model codes
Catalog:110003173, Lot 127480, 185420, 233500, 233520
510(k) numbers
["K123485"]
PMA numbers
not on file
Product code
MBI
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register