Varian Medical Systems, Inc. Oncology Systems: TrueBeam and TrueBeam STx, Model Number: H19, Ref/FSCA identifier: CP-06381 are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
- Recall number
- Z-0364-2012
- Recalling firm
- Varian Medical Systems, Inc. Oncology Systems
- Product
- TrueBeam and TrueBeam STx, Model Number: H19, Ref/FSCA identifier: CP-06381 are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
- Status
- Terminated
- Initiated
- 2011-10-31
- Posted
- 2011-12-02
- Terminated
- 2012-07-25
- Reason
- An anomaly has been identified with the TrueBeam and TrueBeam STx systems where, under certain tuning conditions, the electron beam emerging from the bend magnet may have an elongated spot shape.
- Root cause
- Device Design
- Action
- On 11/04/2011, an Urgent Medical Device Correction Letter was distributed to all affected users, with a description of the problem and user corrective action steps. Varian is developing a corrective action for this issue. A customer service representative will contact you when the correction is available to schedule installation on your system. Questions regarding the letter should be directed to the Varian Oncology Help Desk Contact Information at USA and Canada: 1-888-827- 4265 or Europe +41-41-749-8844.
- Quantity
- not on file
- Distribution
- Worldwide distribution including the US and Internationally.
- Lot, serial or model codes
- Model numbers: H191113, H191034, H191105, H191154, H191046, H191053, H191044, H191015, H191035, H191108, 191152, H191140, H191069, H191017, H191027, H191109, 191130, H191002, H191071, H191024, H191078, H191111, 191072, H191102, H191074, H191087, H191151, H191125, 191018, H191058, H191038, H191115, H191085, H191032, 191077, H191023, H191103, H191095, H191163, H191146, 191124, H191138, H191093, H191097, H191040, H191030, 191029, H191060, H191066, H191122, H191054, H191126, 191114, H191082, H191067, H191081, H191005, H191013, 191116, H191051, H191094, H191052, H191031, H191021, 191048, H191139, H191099, H191020, H191039, H191128, 191156, H191057, H191061, H191112, H191049, H191083, 191008, H191123, H191065, H191119, H191089, H191120, 191092, H191134, H191129, H191056, H191090, H191033, 191047, H191037, H191076, H191007, H191028, H191080, 191106, H191118, H191073, H191117, H191155, H191086, 191141, H191025, H191036, H191042, H191079, H191135, 191010, H191100, H191107, H191127, H191063, H191070, 191153, H191026, H191096, H191014, H191064, H191059, 191009, H191142, H191068, H191001, H191132, H191011, 191003, H191136, H191022, H191121, H191016, H191012, 191045, H191101, H191088, H191041, H191043, H191150, 191019, H191055, H191133, H191131, H191091, H191098, 191006 and H191110
- 510(k) numbers
- ["K111106"]
- PMA numbers
- not on file
- Product code
- -IYE
- Firm FEI
- 2916710
- Firm city
- Palo Alto
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28