Varian Medical Systems, Inc. Oncology Systems: TrueBeam and TrueBeam STx, Model Number: H19, Ref/FSCA identifier: CP-06381 are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.

Recall number
Z-0364-2012
Recalling firm
Varian Medical Systems, Inc. Oncology Systems
Product
TrueBeam and TrueBeam STx, Model Number: H19, Ref/FSCA identifier: CP-06381 are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
Status
Terminated
Initiated
2011-10-31
Posted
2011-12-02
Terminated
2012-07-25
Reason
An anomaly has been identified with the TrueBeam and TrueBeam STx systems where, under certain tuning conditions, the electron beam emerging from the bend magnet may have an elongated spot shape.
Root cause
Device Design
Action
On 11/04/2011, an Urgent Medical Device Correction Letter was distributed to all affected users, with a description of the problem and user corrective action steps. Varian is developing a corrective action for this issue. A customer service representative will contact you when the correction is available to schedule installation on your system. Questions regarding the letter should be directed to the Varian Oncology Help Desk Contact Information at USA and Canada: 1-888-827- 4265 or Europe +41-41-749-8844.
Quantity
not on file
Distribution
Worldwide distribution including the US and Internationally.
Lot, serial or model codes
Model numbers: H191113, H191034, H191105, H191154, H191046, H191053, H191044, H191015, H191035, H191108, 191152, H191140, H191069, H191017, H191027, H191109, 191130, H191002, H191071, H191024, H191078, H191111, 191072, H191102, H191074, H191087, H191151, H191125, 191018, H191058, H191038, H191115, H191085, H191032, 191077, H191023, H191103, H191095, H191163, H191146, 191124, H191138, H191093, H191097, H191040, H191030, 191029, H191060, H191066, H191122, H191054, H191126, 191114, H191082, H191067, H191081, H191005, H191013, 191116, H191051, H191094, H191052, H191031, H191021, 191048, H191139, H191099, H191020, H191039, H191128, 191156, H191057, H191061, H191112, H191049, H191083, 191008, H191123, H191065, H191119, H191089, H191120, 191092, H191134, H191129, H191056, H191090, H191033, 191047, H191037, H191076, H191007, H191028, H191080, 191106, H191118, H191073, H191117, H191155, H191086, 191141, H191025, H191036, H191042, H191079, H191135, 191010, H191100, H191107, H191127, H191063, H191070, 191153, H191026, H191096, H191014, H191064, H191059, 191009, H191142, H191068, H191001, H191132, H191011, 191003, H191136, H191022, H191121, H191016, H191012, 191045, H191101, H191088, H191041, H191043, H191150, 191019, H191055, H191133, H191131, H191091, H191098, 191006 and H191110
510(k) numbers
["K111106"]
PMA numbers
not on file
Product code
-IYE
Firm FEI
2916710
Firm city
Palo Alto
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register