Skip to the content

FDA Medical Device Recalls (openFDA)

Karl Storz Endoscopy: Irrigation Tubing for KARL STORZ Endomat Select UP210 REF 031523-10 Tubing Set, Irrigation, PC

Recall number
Z-0363-2025
Recalling firm
Karl Storz Endoscopy
Product
Irrigation Tubing for KARL STORZ Endomat Select UP210 REF 031523-10 Tubing Set, Irrigation, PC
Status
Open, Classified
Initiated
2024-10-21
Posted
2024-11-12
Terminated
not on file
Reason
Due to improper labeling of products. An intended use on the label has not been reviewed and approved by FDA for distribution in the U.S.
Root cause
Labeling design
Action
On 10/22/2024, an "URGENT: MEDICAL DEVICE CORRECTION" Letters will be sent via email by KARL STORZ informing customers that the firm has become aware that one lot number of misbranded product was distributed to 19 customers. Customers are instructed to: 1. Pass on this Urgent Medical Device Correction to all users of the products listed above (Table 1) and all other persons who need to be aware within your organization. 2. Communicate to all applicable users that the product is not to be utilized during spinal procedures. 3. If you have or may have distributed the products listed, please identify and promptly notify those recipients of this notice. 4. Return the completed Customer Reply Form by E-Mail to the indicated contact below within 15 calendar days from the date of receipt. 5. Please report any incidents related to this issue to the manufacturer, dealer, or local representative and, if applicable, to the national competent authority, as this is important feedback. for questions send an email to KSEAProductLabeling@karlstorz.com .
Quantity
40 units
Distribution
U.S. Nationwide distribution in the states of CO, CT, DC, MD, MI, MN, NC, NY, OH, OK, TN, TX, WA.
Lot, serial or model codes
Product/Material Number: 031523-10 UDI-DI code: 04048438006306 Lot Number: 240222
510(k) numbers
not on file
PMA numbers
not on file
Product code
NVT
Firm FEI
3007475226
Firm city
El Segundo
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register