Skip to the content

FDA Medical Device Recalls (openFDA)

Bock,Otto,Orthopedic Ind,Inc: Electronic Otto Bock Compact leg prosthesis System, 3C95/3C85, to be used exclusively for the exoprosthetic fitting of amputations of the lower limb. Microprocessor-controlled prosthetic knee joint used by lower extremity amputees.

Recall number
Z-0363-2007
Recalling firm
Bock,Otto,Orthopedic Ind,Inc
Product
Electronic Otto Bock Compact leg prosthesis System, 3C95/3C85, to be used exclusively for the exoprosthetic fitting of amputations of the lower limb. Microprocessor-controlled prosthetic knee joint used by lower extremity amputees.
Status
Terminated
Initiated
2006-08-10
Posted
2007-01-25
Terminated
2008-11-26
Reason
A correction of the Otto Bock Compact prosthetic knee joint, material numbers 3C93 and 3C85, is being conducted. A limited number of Compacts have a defective solder connection between the contact wires and battery terminals. If exposed to strong vibrations these Compact knee joints have an increased probability of malfunction and a resultant risk to the Compact wearer.
Root cause
Other
Action
An Urgent Correction Notice, dated August 11, 2006, was sent to customers who purchased the Compact Prosthetic knee joint. The letter identifies affected units, asks for all in stock devices to be returned to Otto Bock, and for patients fitted with an affected device, a loaner knee will be provided as a temporary replacement while their device is serviced.
Quantity
176
Distribution
Nationwide.
Lot, serial or model codes
Material number 3C85, serial numbers: 200547006, 200547007, 200547008, 200547009, 200547034, 200547035, 200547036, 200547037, 200547038, 200547039, 200548022, 200548023, 200548037, 200548038, 200548039, 200548040, 200548041, 200548042, 200604015, 200604016, 200604017, 200604021, 200604022, 200604024, 200604064, 200605002, 200613003.   Material number 3C95, serial numbers: 200531057, 200531058, 200531059, 200531060, 200547003, 200547004, 200547005, 200547024, 200547049, 200547050, 200547051, 200547052, 200547053, 200547054, 200547055, 200547056, 200547057, 200548007, 200548008, 200548009, 200548010, 200548011, 200548012, 200548013, 200548014, 200548015, 200548016, 200548017, 200548018, 200548019, 200548020, 200548034, 200548035, 200548036, 200548044, 200548045, 200548046, 200548047, 200548048, 200548049, 200548051, 200548052, 200548053, 200548054, 200548055, 200548056, 200548058, 200548059, 200548060, 200548061, 200548062, 200548063, 200548064, 200548065, 200548067, 200548068, 200604001, 200604002, 200604003, 200604004, 200604005, 200604006, 200604007, 200604008, 200604009, 200604010, 200604012, 200604013, 200604014, 200604026, 200604027, 200604028, 200604029, 200604030, 200604032, 200604033, 200604034, 200604035, 200604036, 200604038, 200604039, 200604040, 200604041, 200604042, 200604043, 200604044, 200604045, 200604053, 200604054, 200604055, 200604056, 200604057, 200604058, 200604059, 200604062, 200604063, 200605003, 200605004, 200605005, 200605006, 200605008, 200605009, 200605010, 200605018, 200605019, 200605020, 200605021, 200605022, 200605023, 200605054, 200605055, 200605057, 200605058, 200605059, 200605060, 200605061, 200612015, 200612016, 200612017, 200612018, 200612019, 200612020, 200612021, 200612024, 200612029, 200612045, 200612047, 200612050, 200612051, 200612052, 200612053, 200612054, 200612056, 200613034, 200613035, 200613038, 200613039, 200613041, 200613042, 200613043, 200613044, 200613045, 200613054, 200613055, 200613056, 200613057, 200613061, 200613062, 200613063.
510(k) numbers
["K991590"]
PMA numbers
not on file
Product code
KFX
Firm FEI
2182293
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register