FDA Medical Device Recalls (openFDA)
Bock,Otto,Orthopedic Ind,Inc: Electronic Otto Bock Compact leg prosthesis System, 3C95/3C85, to be used exclusively for the exoprosthetic fitting of amputations of the lower limb. Microprocessor-controlled prosthetic knee joint used by lower extremity amputees.
- Recall number
- Z-0363-2007
- Recalling firm
- Bock,Otto,Orthopedic Ind,Inc
- Product
- Electronic Otto Bock Compact leg prosthesis System, 3C95/3C85, to be used exclusively for the exoprosthetic fitting of amputations of the lower limb. Microprocessor-controlled prosthetic knee joint used by lower extremity amputees.
- Status
- Terminated
- Initiated
- 2006-08-10
- Posted
- 2007-01-25
- Terminated
- 2008-11-26
- Reason
- A correction of the Otto Bock Compact prosthetic knee joint, material numbers 3C93 and 3C85, is being conducted. A limited number of Compacts have a defective solder connection between the contact wires and battery terminals. If exposed to strong vibrations these Compact knee joints have an increased probability of malfunction and a resultant risk to the Compact wearer.
- Root cause
- Other
- Action
- An Urgent Correction Notice, dated August 11, 2006, was sent to customers who purchased the Compact Prosthetic knee joint. The letter identifies affected units, asks for all in stock devices to be returned to Otto Bock, and for patients fitted with an affected device, a loaner knee will be provided as a temporary replacement while their device is serviced.
- Quantity
- 176
- Distribution
- Nationwide.
- Lot, serial or model codes
- Material number 3C85, serial numbers: 200547006, 200547007, 200547008, 200547009, 200547034, 200547035, 200547036, 200547037, 200547038, 200547039, 200548022, 200548023, 200548037, 200548038, 200548039, 200548040, 200548041, 200548042, 200604015, 200604016, 200604017, 200604021, 200604022, 200604024, 200604064, 200605002, 200613003. Material number 3C95, serial numbers: 200531057, 200531058, 200531059, 200531060, 200547003, 200547004, 200547005, 200547024, 200547049, 200547050, 200547051, 200547052, 200547053, 200547054, 200547055, 200547056, 200547057, 200548007, 200548008, 200548009, 200548010, 200548011, 200548012, 200548013, 200548014, 200548015, 200548016, 200548017, 200548018, 200548019, 200548020, 200548034, 200548035, 200548036, 200548044, 200548045, 200548046, 200548047, 200548048, 200548049, 200548051, 200548052, 200548053, 200548054, 200548055, 200548056, 200548058, 200548059, 200548060, 200548061, 200548062, 200548063, 200548064, 200548065, 200548067, 200548068, 200604001, 200604002, 200604003, 200604004, 200604005, 200604006, 200604007, 200604008, 200604009, 200604010, 200604012, 200604013, 200604014, 200604026, 200604027, 200604028, 200604029, 200604030, 200604032, 200604033, 200604034, 200604035, 200604036, 200604038, 200604039, 200604040, 200604041, 200604042, 200604043, 200604044, 200604045, 200604053, 200604054, 200604055, 200604056, 200604057, 200604058, 200604059, 200604062, 200604063, 200605003, 200605004, 200605005, 200605006, 200605008, 200605009, 200605010, 200605018, 200605019, 200605020, 200605021, 200605022, 200605023, 200605054, 200605055, 200605057, 200605058, 200605059, 200605060, 200605061, 200612015, 200612016, 200612017, 200612018, 200612019, 200612020, 200612021, 200612024, 200612029, 200612045, 200612047, 200612050, 200612051, 200612052, 200612053, 200612054, 200612056, 200613034, 200613035, 200613038, 200613039, 200613041, 200613042, 200613043, 200613044, 200613045, 200613054, 200613055, 200613056, 200613057, 200613061, 200613062, 200613063.
- 510(k) numbers
- ["K991590"]
- PMA numbers
- not on file
- Product code
- KFX
- Firm FEI
- 2182293
- Firm city
- Minneapolis
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28