FDA Medical Device Recalls (openFDA)

Ossur North America Inc: Fixed Offset Adapter, A-642100. The Offset Adapter is a fixed modular component which sets the tube 14.5 mm (0.57 inch) offset from the center line of the pyramid. It is made of a high-strength alloy with a pyramid receiver and is intended where there is a need to offset the tube horizontally in relation to other components. It is fully compatible with all standard 30mm prosthetic pylons. (3) Offset: 14,5mm (0.57inch).

Recall number
Z-0362-2014
Recalling firm
Ossur North America Inc
Product
Fixed Offset Adapter, A-642100. The Offset Adapter is a fixed modular component which sets the tube 14.5 mm (0.57 inch) offset from the center line of the pyramid. It is made of a high-strength alloy with a pyramid receiver and is intended where there is a need to offset the tube horizontally in relation to other components. It is fully compatible with all standard 30mm prosthetic pylons. (3) Offset: 14,5mm (0.57inch).
Status
Terminated
Initiated
2013-10-24
Posted
2013-11-19
Terminated
2014-06-17
Reason
Ossur inititated a voluntary recall of the A-642100 Fixed Offset Adapter because it can develop fatigue cracks in normal use over an extended period of time and ultimately separate completely.
Root cause
Device Design
Action
Ossur sent an "Safety Notice Voluntary Product Recall" letter dated October 29, 2013, via regular mail. Follow-up telephones are also made. The letter identified the product the problem and the actions to be taken by the customer. ssur is committed to providing superior customer service and we will do our utmost to minimize any inconvenience this may cause your practice. To ensure patient safety we kindly ask you to take the following action: " Please identify the users of each of the A-642100 adapters, see attached list. " Also, please check your inventory for any A-642100s you may have on hand but have not fitted on users. A follow-up phone call will be made a few days later to coordinate replacements. A new replacement component will be provided to the customer. For further questions please call (949) 360-3634.
Quantity
66 units
Distribution
US Distribution including the states of AL, TX, NY, CA, VA, GA, WI, NC, ID and NJ.
Lot, serial or model codes
Item Number: A-642100    All Serial Numbers.
510(k) numbers
not on file
PMA numbers
not on file
Product code
ISH
Firm FEI
1000519909
Firm city
Foothill Ranch
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register