FDA Medical Device Recalls (openFDA)
SCC Soft Computer: SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7. SoftReports is a web-based report launching tool to be used in a hospital or laboratory environment by knowledgeable, trained, and experienced personnel. SoftReports enables the user to schedule and generate reports for the purpose of viewing and printing data stored in SCC product databases.
- Recall number
- Z-0362-2012
- Recalling firm
- SCC Soft Computer
- Product
- SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7. SoftReports is a web-based report launching tool to be used in a hospital or laboratory environment by knowledgeable, trained, and experienced personnel. SoftReports enables the user to schedule and generate reports for the purpose of viewing and printing data stored in SCC product databases.
- Status
- Terminated
- Initiated
- 2011-10-25
- Posted
- 2011-12-02
- Terminated
- 2012-03-16
- Reason
- On 10/25/2011 SCC Soft Computer, Clearwater, FL initiated a correction on the SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7. When SoftReports Designer opened a layout, the system did not load the layout with the previously saved value in the NewPage property. This caused multiple patient records to be printed on the same page. All affected clients were notified on the issue on
- Root cause
- Software design
- Action
- SCC Soft Computer sent a "Correction Notification" letter dated October 25, 2011 to all affected customers. The letter described the product, problem, and actions to be taken by the customers. The hot fixes that were previously loaded which introduced this issue were corrected and delivered to all affected customers as a mandatory correction in hot fixes 1.7199.8 and 1.7200.8 and higher.
- Quantity
- 5
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LNX
- Firm FEI
- 1000307651
- Firm city
- Clearwater
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28