Skip to the content

FDA Medical Device Recalls (openFDA)

Avanos Medical, Inc.: AVANOS BALLARD Closed Suction System for Adults, 14 Fr, DSE, MDI, Product code 221038, provide suction on patients on mechanical ventilation

Recall number
Z-0360-2024
Recalling firm
Avanos Medical, Inc.
Product
AVANOS BALLARD Closed Suction System for Adults, 14 Fr, DSE, MDI, Product code 221038, provide suction on patients on mechanical ventilation
Status
Open, Classified
Initiated
2023-10-13
Posted
2023-11-21
Terminated
not on file
Reason
Certain lots of BALLARD Closed Suction System for Adults, 14 Fr, DSE, MDI were labeled as not being made with Di(2-ethylhexyl) phthalate (DEHP); however, Avanos has determined that the tubing components of the product did contain DEHP.
Root cause
Labeling Change Control
Action
Avanos medical issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 10/13/2023 via FedEx. The notice explained the issue, potential risk to health, and requested that the product be destroyed at the consignee. Distributors were directed to notify their customers.
Quantity
4730 cases (94600 devices)
Distribution
US and South Africa
Lot, serial or model codes
UDI/DI 00609038945064, Lot Numbers: 30112716, 30122936, 30116495, 30122179, 30116496, 30124942, 30117487, 30126227, 30117488, 30124946, 30117632, 30127429, 30117633, 30127494, 30113925, 30137504, 30119101, 30153170, 30119102, 30151042, 30116223, 30194152, 30123991
510(k) numbers
not on file
PMA numbers
not on file
Product code
BSY
Firm FEI
3011270181
Firm city
Alpharetta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register