Vapotherm, Inc.: Vapotherm" Respiratory Gas Humidifier, Model 2000i, labeled in part ***Vapotherm Inc., Annapolis, MD 21401***.

Recall number
Z-0360-06
Recalling firm
Vapotherm, Inc.
Product
Vapotherm" Respiratory Gas Humidifier, Model 2000i, labeled in part ***Vapotherm Inc., Annapolis, MD 21401***.
Status
Terminated
Initiated
2005-10-13
Posted
2006-01-19
Terminated
2008-08-28
Reason
A medical device firm changed the labeling of its respiratory gas humidifier by revising the protocol for higher level disinfectants to be used by healthcare practitioners.
Root cause
Other
Action
Vapotherm, Inc. reportedly notified distributors and institutional end users via subscriber list-serve with newsletters sent on 10/13/05, 10/31/05 and 11/04/05. The notifications addressed the CDC epidemiological study, in which no definitive information on the source of the organism was identified. The notice states that the firm continued to recommend disinfection between patients and adherence to FDA limitations on the re-use of single patient consumables. Subsequently, Vapotherm reported that they had developed an enhanced disinfecting protocol including a recommendation for using higher level disinfectants and the use of sterile water in the system. The notification refers consignees to the company website for infection prevention and control resources updates. On 12/20/05, Vapotherm notified their list-serve subscribers that they were recalling 2000i devices to conduct a high level disinfectant protocol with the devices to be returned to end users with newly revised labeling to include recommendation for sterile water use and device components designed to provide humidification within a closed system. On 12/22/05, Vapotherm notified distributors and healthcare practitioners via certified mail flagged 'Urgent'. The firm's notification was reportedly sent to hospital respiratory and infection control departments and advised that other therapies, such as nasal CPAP and standard oxygen therapies should be considered during the recall. The notification directs users to the firm's website where users may log on and obtain return authorization for shipping back to the recalling firm. A response form is included with the notification and users are requested to fax back for confirmation of notification.
Quantity
5092 units
Distribution
Products were sold to medical device distributors and end users nationwide and in international markets.
Lot, serial or model codes
All S/Ns of the Vapotherm" Model 2000i are affected.
510(k) numbers
not on file
PMA numbers
not on file
Product code
BTT
Firm FEI
3002877041
Firm city
Stevensville
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register