Skip to the content

FDA Medical Device Recalls (openFDA)

Ebi, Llc: BIOMET Spine, 5.5 TI 5.5x50mm Translation Screw Polaris 5.5 TI Spinal System. Rx Only. Non-Sterile. Intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system

Recall number
Z-0359-2014
Recalling firm
Ebi, Llc
Product
BIOMET Spine, 5.5 TI 5.5x50mm Translation Screw Polaris 5.5 TI Spinal System. Rx Only. Non-Sterile. Intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system
Status
Terminated
Initiated
2013-08-22
Posted
2013-11-19
Terminated
2016-08-31
Reason
Polaris Translation Screw Failures were recently reported to Biomet Spine. These failures are generally described as dissociation of the screw head from the shaft of the Polaris Translation Screw.
Root cause
Device Design
Action
Urgent Product Information notices dated 8/22/2013 were mailed via FedEx to distributors on 8/23/2013. Follow-up emails with a copy of the UPI notice were sent to distributors on 8/23/2013. The UPI notice informed the customers of the issue with the product and how to identify the affected product. Customers were advised that the Polaris Translation Screw System and the SpF Implantable Spine Fusion Stimulator not be used in conjunction with any implants containing cobalt chrome. Firm contacts were provided in case customers had any questions. EBI, LLC had a conference call with their domestic distributors on 8/23/2013 informing them they are required to notify their customers.
Quantity
7,092 units
Distribution
Nationwide distribution including Puerto Rico.
Lot, serial or model codes
Multiple Item and Lot numbers. The firm expects all Translation Screw Systems (containing a cobalt chrome component) used and contacted with the  SpF Implantable Spine Fusion Stimulator (DC electrical current) to be at risk for this failure mode.
510(k) numbers
["K123549"]
PMA numbers
not on file
Product code
HWC
Firm FEI
2242816
Firm city
Parsippany
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register