FDA Medical Device Recalls (openFDA)
Boston Scientific Corporation: Boston Scientific brand Renegade Fiber Braided Microcatheter, 150/10/2RO, Rx Only; Model Numbers: 18257, 18259; Product is distributed by Boston Scientific Corporation, Natick, MA; Product is manufactured by Boston Scientific Corporation, Cork, Ireland
- Recall number
- Z-0358-2011
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific brand Renegade Fiber Braided Microcatheter, 150/10/2RO, Rx Only; Model Numbers: 18257, 18259; Product is distributed by Boston Scientific Corporation, Natick, MA; Product is manufactured by Boston Scientific Corporation, Cork, Ireland
- Status
- Terminated
- Initiated
- 2009-08-24
- Posted
- 2010-11-15
- Terminated
- 2011-02-03
- Reason
- Sterility Loss-- The sterile barrier pouch seal (top seal) may be breached by the product, resulting in loss of sterility which may result in transmission of disease or infection.
- Root cause
- Packaging process control
- Action
- Boston Scientific issued an Urgent Medical Device Recall letter dated August 26, 2009 identifying the affected devices, the issue prompting the recall, and actions to be taken by customers. Customers were instructed to determine if they have affected product in their inventory, and if so, immediately quarantine and return to Boston Scientific. Customers can contact Boston Scientific at 510 624-1734.
- Quantity
- not on file
- Distribution
- Distribution Worldwide, including throughout the US.
- Lot, serial or model codes
- Lots: 11249879, 11542408, 11750295, 11966695, 12156146, 12228393, 12365497, 12491154, 11253574, 11546053, 11756331, 11988066, 12156147, 12228443, 12365604, 12509751, 11257185, 11552115, 11776786, 11996448, 12156183, 12228444, 12365605, 12509758, 11268801, 11554501, 11793108, 12001306, 12170507, 12248055, 12365613, 12509920, 11271408, 11559603, 11800162, 12001562, 12170508, 12248109, 12365614, 12527317, 11562655, 11274973, 11572313, 11806766, 12001567, 12170516, 12248121, 12385142, 12527760, 11296064, 11573180, 11819657, 12001568, 12175704, 12262088, 12385143, 12527763, 11301389, 11577242, 11825864, 12021328, 12175706, 12270709, 12385158, 12527769, 11303513, 11580376, 11836834, 12028753, 12175708, 12270714, 12385224, 12546795, 11840079, 11307943, 11586443, 11843094, 12039033, 12175710, 12270983, 12402095, 12546199, 11318319, 11587960, 11849818, 12049925, 12175711, 12279662, 12402111, 12546941, 11319596, 11602048, 11857667, 12061129, 12175713, 12292373, 12421265, 12546946, 11423369, 11623809, 11869009 12068930, 12175717, 12296121, 12421267, 12560791, 12073876, 11425310, 11627605, 11876193, 12089556, 12175719, 12296833, 12421271, 12560192 11425314, 11631734, 11880345, 12104631, 12175724, 12297455, 12421274, 12578834 11429065, 11638177, 11884227, 12104678, 12194075, 12311446, 12439946, 12578835 11429069, 11649446, 11891449, 12108064, 12194077, 12311459, 12439951, 12578864 12194829, 11432583, 11656601, 11891458, 12120518, 12194831, 12311520, 12439954, 12614498, 11432584, 11678955, 11895180, 12120528, 12214586, 12326062, 12439958, 12634309, 11441965, 11682874, 11895183, 12123141, 12214589, 12330440, 12458414, 12634310, 11444027, 11695922, 11925162, 12133525, 12214592, 12330441, 12458475, 12634315, 12330443, 11445708, 11703970, 11927490, 12142142, 12214602, 12330465, 12458416, 12659579, 11461698, 11706368, 11933896, 12142150, 12214604, 12330476, 12458543, 12660000, 11469070, 11710717, 11941913, 12148683, 12214616, 12330511, 12458548, 11481564, 11714291, 11948975, 12148684, 12228116, 12348609, 12466721, 11534917, 11721375, 11959863, 12148687, 12228366, 12365495, 12466123,
- 510(k) numbers
- ["K973645"]
- PMA numbers
- not on file
- Product code
- KRA
- Firm FEI
- 2939204
- Firm city
- Fremont
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28