Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems North America Co. Phillips: Gyroscan NT MRI System rev 4.x, 5.x, and 6.x software. System, Nuclear Magnetic Resonance imaging.

Recall number
Z-0358-06
Recalling firm
Philips Medical Systems North America Co. Phillips
Product
Gyroscan NT MRI System rev 4.x, 5.x, and 6.x software. System, Nuclear Magnetic Resonance imaging.
Status
Terminated
Initiated
2005-12-09
Posted
2006-01-06
Terminated
2008-05-20
Reason
The Gyroscan NT MRI system has a computer software problem which results in images from one patient being placed into another patient's image record.
Root cause
Other
Action
Product safety notification sent certified mail to each customer explaining the problem. A mandatory software upgrade to correct the problem will be installed starting the first quarter of 2006 and be completed by the third quarter of 2006.
Quantity
96
Distribution
96 units distributed to 92 U.S. Customers
Lot, serial or model codes
4397, 5257, 4176, 1052, 4222, 4064, 5024, 4009, 4241, 5313, 1139, 5484, 4328, 5201,  5246, 5381, 4296, 4258, 5277, 4246, 1240, 5297, 5311, 5434, 5512, 506, 4207, 1072, 4089, 5373, 1014, 1078, 4025, 5551, 4131, 5455, 5328, 5379, 5261, 4416, 4206, 1191, 4088, 5195, 5334, 5451, 4383, 5250, 5309, 5342, 4183, 4272, 5268, 5009, 4047, 4254,  4261, 5307, 4102, 5199, 5558, 5315, 5266, 5303, 1061, 4307, 5390, 4267, 4386, 5282,  1113, 5123, 5501, 1062, 4339, 5383, 5478, 5507, 4162, 1015, 5162, 1081, 4170, 5477,  4322, 5515, 4069, 5081, 4201, 354, 5345, 4374, 4146, 1097, 5519, 5314
510(k) numbers
["K940534"]
PMA numbers
not on file
Product code
LNH
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register