FDA Medical Device Recalls (openFDA)
Cardinal Health: Dover Closed Urethral Tray with Hydrogel Coated Red Rubber Catheter 16 Fr/Ch (5.3 mm) Product Code: 3408 Intended for urinary drainage from the bladder of a patient
- Recall number
- Z-0357-2023
- Recalling firm
- Cardinal Health
- Product
- Dover Closed Urethral Tray with Hydrogel Coated Red Rubber Catheter 16 Fr/Ch (5.3 mm) Product Code: 3408 Intended for urinary drainage from the bladder of a patient
- Status
- Open, Classified
- Initiated
- 2022-11-01
- Posted
- 2022-12-02
- Terminated
- not on file
- Reason
- Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of urinary retention
- Root cause
- Equipment maintenance
- Action
- Cardinal Health issued Urgent Medical Device Recall Letter (Event-2022-04944) and Acknowledgement Form on November 1, 2022 to the direct consignees. Letter states reason for recall, health risk and action to take: Actions Required: 1. REVIEW your inventory for the affected product codes and lots. Location of product codes and lots are shown in Attachment 1. 2. SEGREGATE and quarantine all affected product upon review of your inventory. 3. NOTIFY any customers to whom you may have distributed/forwarded affected product to or will send the product on to about this product recall notice and share a copy of this notice. 4. RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to GMB-FieldCorrectiveAction@cardinalhealth.com, whether you have affected product. CONTACT the appropriate Customer Service group to arrange for return and credit/replacement of any affected product Monday Friday between 8:00am - 5pm EST: " Hospital 800-965-5227 " Federal Government 800-444-1166 " Distributor 800-635-6021 " All Other Customers 888-444-5440 For questions related to this notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: GMB-FieldCorrectiveAction@cardinalhealth.com or call 800-292-9332.
- Quantity
- not on file
- Distribution
- Nationwide
- Lot, serial or model codes
- UDI-DI: Each 10884521009431 Case 20884521009438 Lot Numbers: 2215825664
- 510(k) numbers
- ["K903340"]
- PMA numbers
- not on file
- Product code
- FCM
- Firm FEI
- 3014524790
- Firm city
- Mansfield
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28