Skip to the content

FDA Medical Device Recalls (openFDA)

Integra LifeSciences Corp.: Integra Surgical Luxtec MLX Light Source Catalog Numbers: 00MLXAU Product Usage: The Integra Luxtec MLX Light Source is designed to supply high intensity white light through a fiber optic cable for illumination of a surgical field during a surgical and or medical procedure.

Recall number
Z-0357-2012
Recalling firm
Integra LifeSciences Corp.
Product
Integra Surgical Luxtec MLX Light Source Catalog Numbers: 00MLXAU Product Usage: The Integra Luxtec MLX Light Source is designed to supply high intensity white light through a fiber optic cable for illumination of a surgical field during a surgical and or medical procedure.
Status
Terminated
Initiated
2011-10-14
Posted
2011-12-02
Terminated
2012-03-08
Reason
A loose screw may result in shifting of components and improper function of the turret or actuator assembly.
Root cause
Process control
Action
Integra sent an URGENT: Field Corrective Action letter dated October 14, 2011 to all affected customers via FedEx. The letter identified the affected product, the problem, and the actions to be taken with instructions to immediately inspect their inventory for potentially loose screws within the affected unit(s). If the unit(s) were observed with loose screws, the unit(s) were to be returned to Integra Surgical for repair and servicing. Customers were instructed to complete the Acknowledgement and Return Form per the instructions on the form. For questions call Integra Surgical Customer Service at 1-800-431-1123.
Quantity
12
Distribution
Worldwide Distribution - USA (nationwide) including the states of: Arizona, California, Colorado, Florida, Indiana, Mississippi, New Jersey, New York, North Carolina, Ohio, Pennsylvania, Montana, Washington, Wyoming, and Puerto Rico, and the countries of: Australia, Germany, Italy, Japan, Netherlands, New Zealand, Pakistan, Sweden, Switzerland, and Turkey.
Lot, serial or model codes
SN'S: P0102211 P0102311 P0104611 P0104811 P0105011 P0105311 P0105411 P0105511 P0105911 P0106511 P0100311
510(k) numbers
["K081477"]
PMA numbers
not on file
Product code
FCW
Firm FEI
3003418325
Firm city
Plainsboro
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register