Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: Software Package, iPAW 4.2 User Kit, Part No. B73489, Manufactured by: Beckman Coulter Biomedical GmbH The iPAW 4.2 User Kit (PN B73489) includes the iPAW software application CD (version 4.2). This function is primarily used to copy an existing barcode ID to replace the label of an individual sample tube because the original label is deteriorated.

Recall number
Z-0356-2019
Recalling firm
Beckman Coulter Inc.
Product
Software Package, iPAW 4.2 User Kit, Part No. B73489, Manufactured by: Beckman Coulter Biomedical GmbH The iPAW 4.2 User Kit (PN B73489) includes the iPAW software application CD (version 4.2). This function is primarily used to copy an existing barcode ID to replace the label of an individual sample tube because the original label is deteriorated.
Status
Terminated
Initiated
2018-05-24
Posted
not on file
Terminated
2022-12-08
Reason
Sorting-Drive software version 4.2 (released in October 2015) introduced a new option for iPAW and the Sorting-Drive Client that is used for manual printing of barcodes. When printing consecutively numbered tube labels using this optional Barcode Counter function, the system incorrectly converts numbers with consecutive multiple zeroes in the middle, potentially leading to wrong barcode IDs. The firm reports that this does not affect automated label creation or reprinting of individual labels and that no other system functions are affected.
Root cause
Software change control
Action
On 6/4/2018 Beckman Coulter (BEC) reported that the issue will be corrected via a field modification performed by Field Service. BEC reported that they do not intend to send a customer letter for this issue.US customers can contact the Technical Support Troubleshooting Hotline at (800) 854-3633.
Quantity
14
Distribution
Worldwide Distribution: US (nationwide) and countries of: Algeria, Argentina, Austria, Belgium, Brazil, Chile, China, Czech Republic, Denmark ,France, Germany, Greece ,Hong Kong, India, Ireland, Israel, Italy ,Republic of Korea, Lebanon, Lithuania ,Malaysia, Netherlands, Poland, Portugal , Qatar, Romania, Russian Federation, Saudi Arabia ,South Africa , Spain, Sweden, Taiwan, Turkey , and United Kingdom.
Lot, serial or model codes
v4.2.1 (and 4.2, discontinued)
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register