Skip to the content

FDA Medical Device Recalls (openFDA)

Restoration Robotics: ARTAS Disposable and Reuseable Kit ( Contains Skin Tensioner FP-45384; Inner Needle with Vacuum Tube FP-45385; dissection Punch 4.0mm FP 23304; Saline Nose Cone FP-45387; Follicle Trap FP 23306; Non-Sterile Disposable kit FP 23267 and Molded Parts, Reusable Kit FP 45538 Rx): Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.

Recall number
Z-0356-2014
Recalling firm
Restoration Robotics
Product
ARTAS Disposable and Reuseable Kit ( Contains Skin Tensioner FP-45384; Inner Needle with Vacuum Tube FP-45385; dissection Punch 4.0mm FP 23304; Saline Nose Cone FP-45387; Follicle Trap FP 23306; Non-Sterile Disposable kit FP 23267 and Molded Parts, Reusable Kit FP 45538 Rx): Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.
Status
Terminated
Initiated
2013-10-13
Posted
2013-11-18
Terminated
2013-12-10
Reason
Reports of discoloration (residue) on the pins of the skin tensioner have been received by several customers. Discoloration due to insufficient cleaning of stainless steel pins during passivation process. The residue has been identified as a compound of iron and chlorine.
Root cause
Process control
Action
Restoration Robotics sent an Urgent Recall letters dated October 13, 2013 to all distributors via first class mail. Direct accounts were notified by phone as well as fax. Customers were informed of the affected product, problem and actions to be taken. Replacement products are being distributed to each site and will be delivered to each customer. For questions call 408-883-6760
Quantity
515 US, 315 International (all Packaging configurations.
Distribution
Worldwide Distribution - USA Nationwide in the countries of: Korea, Japan, Singapore, Thailand, United Kingdom, Canada, and Taiwan.
Lot, serial or model codes
Model number FP-45399
510(k) numbers
["K123548"]
PMA numbers
not on file
Product code
ONA
Firm FEI
3006515340
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register