FDA Medical Device Recalls (openFDA)
Restoration Robotics: ARTAS Disposable and Reuseable Kit ( Contains Skin Tensioner FP-45384; Inner Needle with Vacuum Tube FP-45385; dissection Punch 4.0mm FP 23304; Saline Nose Cone FP-45387; Follicle Trap FP 23306; Non-Sterile Disposable kit FP 23267 and Molded Parts, Reusable Kit FP 45538 Rx): Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.
- Recall number
- Z-0356-2014
- Recalling firm
- Restoration Robotics
- Product
- ARTAS Disposable and Reuseable Kit ( Contains Skin Tensioner FP-45384; Inner Needle with Vacuum Tube FP-45385; dissection Punch 4.0mm FP 23304; Saline Nose Cone FP-45387; Follicle Trap FP 23306; Non-Sterile Disposable kit FP 23267 and Molded Parts, Reusable Kit FP 45538 Rx): Product Usage: Skin Tensioner is applied to patients scalp during follicular harvesting during hair restoration procedure. Product applies tension to scalp to facilitate follicular extractions.
- Status
- Terminated
- Initiated
- 2013-10-13
- Posted
- 2013-11-18
- Terminated
- 2013-12-10
- Reason
- Reports of discoloration (residue) on the pins of the skin tensioner have been received by several customers. Discoloration due to insufficient cleaning of stainless steel pins during passivation process. The residue has been identified as a compound of iron and chlorine.
- Root cause
- Process control
- Action
- Restoration Robotics sent an Urgent Recall letters dated October 13, 2013 to all distributors via first class mail. Direct accounts were notified by phone as well as fax. Customers were informed of the affected product, problem and actions to be taken. Replacement products are being distributed to each site and will be delivered to each customer. For questions call 408-883-6760
- Quantity
- 515 US, 315 International (all Packaging configurations.
- Distribution
- Worldwide Distribution - USA Nationwide in the countries of: Korea, Japan, Singapore, Thailand, United Kingdom, Canada, and Taiwan.
- Lot, serial or model codes
- Model number FP-45399
- 510(k) numbers
- ["K123548"]
- PMA numbers
- not on file
- Product code
- ONA
- Firm FEI
- 3006515340
- Firm city
- San Jose
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28