Skip to the content

FDA Medical Device Recalls (openFDA)

NuVasive Inc: 6.5mm x 30mm Cannulated Screw Shank Subsequent Codes: KWP, KWQ, MNH, MNI

Recall number
Z-0355-2013
Recalling firm
NuVasive Inc
Product
6.5mm x 30mm Cannulated Screw Shank Subsequent Codes: KWP, KWQ, MNH, MNI
Status
Terminated
Initiated
2009-01-05
Posted
2012-11-19
Terminated
2012-11-19
Reason
The tulip portion of the screw assembly can disengage from the shank after implantation.
Root cause
Device Design
Action
Telephone communications to the sales representatives of surgeons actively using the Posted Pedicle Screw System were made on 01/05/2009. The calls were unscripted and requested that all inventory be forwarded to NuVasive's San Diego office for quarantine. Surgeons who were actively using the system were also contacted and notified of the situation. Follow-up communications were sent via mail on 01/19/2009 to all surgeons who participated in the Alpha release. The letter informed the customers of the problems identified. Customers were instructed that the recalling firm does not believe any further action is required and they were instructed to monitor the progress of their patients and notify NuVasive if any issues are detected. Customers with questions or concerns are instructed to contact the firm at (858) 909-1964.
Quantity
240 units total
Distribution
Nationwide Distribution-including the states of OR, MO, FL, WI, CA, and NC.
Lot, serial or model codes
Batch: JK2057, JK2157.
510(k) numbers
["K082332"]
PMA numbers
not on file
Product code
NKB
Firm FEI
3002980729
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register