FDA Medical Device Recalls (openFDA)
Medtronic Sofamor Danek USA Inc: Medtronic KYPHON(R) Express(TM) Directional Bone Filler Device, SIZE 2, REF FB, QTY 1 EA, Rx only STERILE R
- Recall number
- Z-0352-2018
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Product
- Medtronic KYPHON(R) Express(TM) Directional Bone Filler Device, SIZE 2, REF FB, QTY 1 EA, Rx only STERILE R
- Status
- Terminated
- Initiated
- 2017-12-06
- Posted
- not on file
- Terminated
- 2019-11-04
- Reason
- Medtronic has discovered the directional arrow at the proximal end of the instrument may not correctly align with the cut out opening on the distal end of the instrument.
- Root cause
- Process control
- Action
- Medtronic distributed a voluntary removal notice via FedEx courier service to impacted U.S. consignees on 06 DEC 2017. The recall notice stated the following: "If he/she has not already done so, your Sales Representative will contact you to facilitate the return of any impacted products you may have in your possession. Please disseminate this information to additional personnel within your facility as appropriate, and maintain a copy of this notice in your records. Instructions to Customers: 1.) Please locate and remove the impacted product from normal storage locations. Do not use this product. 2.) Complete the enclosed Customer Confirmation Form and return 3.) As applicable, contact your Medtronic Sales Representative to coordinate the return and replacement of affected product."
- Quantity
- 3,319 units
- Distribution
- US. Austria, Belgium, Bulgaria, Canada, France, Germany, Italy, Lebanon, Luxembourg, Portugal, South Africa, US
- Lot, serial or model codes
- UDI 00643169097254 Lot Numbers: WI424428, WI428822, WI435227, WI439940, WI442555, WI446722, WI449819, WI455595, WI459477, WI463434, WI472175, WI472178, WI472176, WI472177, WI474575
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- OAR
- Firm FEI
- 1000477302
- Firm city
- Memphis
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28