Skip to the content

FDA Medical Device Recalls (openFDA)

Primus Corporation dba Trinity Biotech: A. MENARINI diagnostics Hb 9210 PREMIER, LEVEL1, LEVEL II, Hemoglobin A1c Controls for hemoglobin A1c Assay, REF 45052, part number 05-01-0926, A1c analysis

Recall number
Z-0349-2023
Recalling firm
Primus Corporation dba Trinity Biotech
Product
A. MENARINI diagnostics Hb 9210 PREMIER, LEVEL1, LEVEL II, Hemoglobin A1c Controls for hemoglobin A1c Assay, REF 45052, part number 05-01-0926, A1c analysis
Status
Terminated
Initiated
2021-10-05
Posted
2022-12-01
Terminated
2023-08-09
Reason
The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. The unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.
Root cause
Nonconforming Material/Component
Action
The firm sent an Urgent Filed Safety Notice their consignees on 10/05/2021 by mail. The notice explained the issue, risk and requested the destruction of of the affected devices.
Quantity
340 kits
Distribution
Turkey, India, Romania, Thailand, Vietnam, S. Korea, Peru, Pakistan Austria, UAE, Chile, Italy.
Lot, serial or model codes
Lot 12161
510(k) numbers
["K952720"]
PMA numbers
not on file
Product code
JPK
Firm FEI
1931251
Firm city
Kansas City
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register