Skip to the content

FDA Medical Device Recalls (openFDA)

Encore Medical, Lp: Foundation Tibial Broach P/N 801-01-013 Foundation Tibial Broach to be used in conjunction with the Tibial Sizing Template and Tibial Broach Guide to prepare the resected tibial plateau for acceptance of a keeled tibial Baseplate stem. P/N 801-01-013

Recall number
Z-0349-2012
Recalling firm
Encore Medical, Lp
Product
Foundation Tibial Broach P/N 801-01-013 Foundation Tibial Broach to be used in conjunction with the Tibial Sizing Template and Tibial Broach Guide to prepare the resected tibial plateau for acceptance of a keeled tibial Baseplate stem. P/N 801-01-013
Status
Terminated
Initiated
2011-10-26
Posted
2011-12-02
Terminated
2012-09-05
Reason
The strike plate may dislodge from the broach stem handle during impaction of the device.
Root cause
Nonconforming Material/Component
Action
DJO Global sent an URGENT FIELD SAFETY NOTICE letter dated October 26, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to pass the information to all those who need to be aware within their organization or to any organization where the affected devices have been transferred. Customers should contact Customer Service to place a replacement order and receive an RMA number and return all affected devices using the RMA number. For any questions call (512) 834-6255.
Quantity
30 units
Distribution
Worldwide Distribution - USA including AR, AZ, CA, LA, MD, NY, TX, UT and Puerto Rico and the countries of United Kingdom and Italy
Lot, serial or model codes
Lot#'s 35149L02, 36204L06, 37990L06, 38272L13, and 50479L12
510(k) numbers
["K970031"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1000116912
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register