Skip to the content

FDA Medical Device Recalls (openFDA)

Greer Laboratories, Inc.: GREER DIA - KIT ANTIGEN M. FAENI, 2.0mL 20 MG/ML, Analyte Specific Reagent, Item: K62, HYPERSENSITIVITY PNEUMONITIS test system, gel diffusion kit for demonstrating precipitating antibodies.

Recall number
Z-0348-2023
Recalling firm
Greer Laboratories, Inc.
Product
GREER DIA - KIT ANTIGEN M. FAENI, 2.0mL 20 MG/ML, Analyte Specific Reagent, Item: K62, HYPERSENSITIVITY PNEUMONITIS test system, gel diffusion kit for demonstrating precipitating antibodies.
Status
Terminated
Initiated
2022-10-25
Posted
2022-11-30
Terminated
2026-02-11
Reason
The product was recalled because stability evaluation of the lots indicated a decrease in reactivity that could lead to a potential false-negative result interpretation. The issue was identified by the firm after failing to meet lot requirements for potency. The lot release error was discovered on October 08, 2021, after a failed potency test was determined during 6-month stability evaluation of retention material for the lots. The subsequent manufacturing investigation, which included a review of the lot release potency documentation, found that the initial lot release potency test for the lot was misinterpreted.
Root cause
Employee error
Action
The firm initiated the recall beginning on 10/25/2022 by telephone and email, following with a "MEDICAL DEVICE RECALL" notice. The notice stated the problem and requested the following: "Greer requests that you immediately discontinue use of the M. faeni Dia-Kit antigen lot(s) and return any remaining stock of the lots in question. Please contact a Greer Customer Service Representative at 1-800-378-3906 between 8 a.m. and 5 p.m. Eastern Time to receive a Returned Goods Authorization number and arrange for return of the product. Your Greer Customer Service Representative will also arrange for replacement of the product."
Quantity
102 units
Distribution
US Distribution to states of: CA, CO, IA, KY, MA, MD, MO, NC, PA, UT, VA, and International to countries of: Canada, Denmark
Lot, serial or model codes
Lot Number 389592, exp 6/12/2023
510(k) numbers
not on file
PMA numbers
not on file
Product code
DGW
Firm FEI
1011574
Firm city
Lenoir
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register