Skip to the content

FDA Medical Device Recalls (openFDA)

Candela Corporation: Candela Laser GentleLase Pro, Powered Laser Surgical Instrument for dermatological use - Model number: 9914-00-9015. and Candela Laser GentleLase Pro LE, Model number: 9914-00-9040

Recall number
Z-0348-2014
Recalling firm
Candela Corporation
Product
Candela Laser GentleLase Pro, Powered Laser Surgical Instrument for dermatological use - Model number: 9914-00-9015. and Candela Laser GentleLase Pro LE, Model number: 9914-00-9040
Status
Terminated
Initiated
2013-11-06
Posted
2013-11-18
Terminated
2015-06-08
Reason
Unintended single pulse maybe emitted to handpiece prior to pressing of trigger or footswitch and cause injury.
Root cause
Software design
Action
Candela notiifed Distributors on 11/8/13 through e-mail, Certified return receipt request, or FedEx package containing distributor letter, customer letter for translation and customer lists. A software update to Version 2.0 will be issued to the user by a field service representative.
Quantity
1258 units
Distribution
Worldwide distribution: US (nationwide) and Internationally to: Azerbaijan, Bulgaria, Chile, Czech Republic, Greece, Israel, Kuwait, Qatar, Saudi Arabia, Serbia, Turkey, and the United Arab Emirates and to Subsidiaries in Australia, France, Germany, Italy, Japan, Portugal, Spain, and the United Kingdom.
Lot, serial or model codes
GL Pro Model 9914-00-9015:: SN 9914-9015-0005 through 9914-9015-1259 GL Pro LE Model 9914-00-9040: SN 9914-9040-0716 through 9914-9040-1214
510(k) numbers
["K024260","K024335","K972767"]
PMA numbers
not on file
Product code
GEX
Firm FEI
1218402
Firm city
Wayland
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register