Skip to the content

FDA Medical Device Recalls (openFDA)

DeRoyal Industries Inc: HYDRO-TEMP(R) Neonatal Temperature Skin Sensor, REF HNICU-32, 50 units per box, NON STERILE, Rx only.

Recall number
Z-0345-2016
Recalling firm
DeRoyal Industries Inc
Product
HYDRO-TEMP(R) Neonatal Temperature Skin Sensor, REF HNICU-32, 50 units per box, NON STERILE, Rx only.
Status
Terminated
Initiated
2015-10-14
Posted
2015-11-30
Terminated
2017-07-01
Reason
DeRoyal manufactured four lots of HNICU-32, Hydro-Temp(R) Neonatal Temperature Skin Sensor incorrectly. The pins contained in the probe were soldered incorrectly. When probe is connected to the proper port, this defect results in the inability to read or display temperatures on the infant warming system.
Root cause
Process control
Action
DeRoyal issued their recall on 10/14/2015 to the end-user level by sending overnight letters to its direct accounts and hospitals. The consignees were instructed to destroy and discard the units. In the case of further distribution, the firm requested that the consignee either notify its customers of the recall or to submit a customer list for DeRoyal to contact them.
Quantity
16 cases of 50 units (800 eaches)
Distribution
Distributed in the states of VA, AL, FL, and CA.
Lot, serial or model codes
Lot Numbers: 39098746, 39655101, 39985370, 40067795
510(k) numbers
["K925006"]
PMA numbers
not on file
Product code
FMT
Firm FEI
1043214
Firm city
Powell
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register