Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics, Inc.: Stratus(R) CS Rotors, catalog number CROTOR, lot number NH42-251-09 Intended use: Stratus(R) rotors is intended for the separation of whole blood samples for analysis on the Stratus(R) CS analyzer

Recall number
Z-0343-2011
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Product
Stratus(R) CS Rotors, catalog number CROTOR, lot number NH42-251-09 Intended use: Stratus(R) rotors is intended for the separation of whole blood samples for analysis on the Stratus(R) CS analyzer
Status
Terminated
Initiated
2009-12-07
Posted
2010-11-10
Terminated
2010-11-10
Reason
Low frequency defect in the gel filling of this lot that may result in inadequate separation of cells in centrifugation.
Root cause
Process control
Action
The recalling firm, SIEMENS, issued an "Urgent Field Safety Notice" dated December 2009 to all affected customers. The notice described the product, problem and the action to be taken by the customers. The customers were instructed to review the situation with their laboratory director to determine the need to evaluate previously reported test results; immediately examine their CROTOR inventory and discard all bags that are from lot NH42-251-09 including rotors that were not stored in their original packaging (bag); contact Siemens Customer Service Center at 800-241-0420 to request additional no-charge replacements (Note: A "no-charge" order has already been placed so that one bag of replacement rotors would be recieved shortly); forward this notice to anyone they may have distributed this product, and complete and return the FIELD CORRECTION EFFECTIVENESS CHECK via fax at 302-631-8467. If you have any questions regarding this information, please contact the Siemens Technical Solution Center at 800-405-6473.
Quantity
330 bags
Distribution
Worldwide distribution: USA including states of AK, AZ, AR, CA, FL, GA, HI, IA, IL, IN, KY, LA, MI, MN, MO, MS, MT, NC, ND, NH, NM, NY, OH, TX, UT, WA, WI, and WV and country of Canada
Lot, serial or model codes
catalog number CROTOR, lot number NH42-251-09
510(k) numbers
["K984067"]
PMA numbers
not on file
Product code
JHX
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register