MRL, Inc., A Welch Allyn Company: Welch Allyn PIC 50 Portable Intensive Care System, Multi-Parameter Monitor/Defibrillator; a portable 12V internal battery powered defibrillator; Manufactured by MRL Inc., a WelchAllyn Company, 1000 Asbury Drive, Buffalo Grove, Illinois 60089-Part Numbers-971081,971082,971083,971084.

Recall number
Z-0343-2007
Recalling firm
MRL, Inc., A Welch Allyn Company
Product
Welch Allyn PIC 50 Portable Intensive Care System, Multi-Parameter Monitor/Defibrillator; a portable 12V internal battery powered defibrillator; Manufactured by MRL Inc., a WelchAllyn Company, 1000 Asbury Drive, Buffalo Grove, Illinois 60089-Part Numbers-971081,971082,971083,971084.
Status
Terminated
Initiated
2006-11-15
Posted
2007-01-19
Terminated
2007-10-31
Reason
Shock Delay:The monitor-defibrillator may display an 'ECG Comm' error message on the display during use, which may prevent or cause an unacceptable delay the delivery of a defibrillation shock. This delay could result in failure to resuscitate the patient.
Root cause
Other
Action
MRL sent Urgent Medical Device Recall letters dated 11/8/06 via certified mail on 11/15/06 to the direct accounts informing them of an intermittent connection between the J5 cable and its mating connector on a circuit board inside of the device that could result in a malfunction that could prevent or unacceptably delay the delivery of a defibrillating shock, which possibly could result in failure to resuscitate the patient. The accounts were provided a list of the affected units'' parts/serial numbers and were requested to inspect their PIC 50 units for the affected serial numbers and return the identified units for servicing. Until their PIC 50 is serviced, the accounts were instructed to check the unit before each shift by using the self-test, which is automatically initiated by powering up the device. If the 'ECG Comm' error occurs during use, cycling the power off-on may clear the error. Even if power cycling clears the device, the accounts were instructed to contact Welch Allyn - MRL to obtain priority servicing of the affected unit. The accounts were also instructed to respond to the notification within 30 days by fax or mail, or by calling Welch Allyn - MRL Technical Support at 1-800-462-0777, to schedule their PIC 50 for preventative service. Welch Allyn will provide the accounts with a loaner PIC 50 after receiving the form with instructions to send back the PIC 50 in the same shipping box accompanying the loaner unit. Any questions were directed to Welch Allyn-MRL Technical Support at 1-800-462-0777.
Quantity
510 units
Distribution
Worldwide-USA, Australia, Belgium, Canada, Egypt, Finland, France, Germany, Ireland, Italy, Jordan, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sultanate of Oman, United Arab Emirates and the United Kingdom.
Lot, serial or model codes
Part Numbers 971081, 971082, 971083 and 971084; Serial numbers 80052, 80141, 80190, 80191, 80425, 80426, 80430, 80431, 80432, 80435, 80436, 80587, 80826, 80915, 80916, 80918, 80920, 80921, 80923, 80924, 80925, 80926, 80927, 80928, 80929, 80930, 80932, 80934, 80960, 80962, 80963, 80967, 80969, 80971, 80972, 80973, 80974, 80975, 80976, 80981, 80983, 80987, 80989, 80990, 80991, 80992, 80993, 81003, 81004, 81005, 81006, 81007, 81008, 81009, 81010, 81011, 81012, 81013, 81014, 81015, 81016, 81017, 81018, 81019, 81020, 81022, 81023, 81024, 81025, 81026, 81027, 81028, 81029, 81030, 81031, 81032, 81033, 81034, 81035, 81036, 81037, 81038, 81044, 81045, 81046, 81047, 81049, 81070, 81094, 81102, 81103, 81104, 81105, 81109, 81110, 81111, 81112, 81113, 81165, 81167, 81178, 81179, 81181, 81184, 81187, 81188, 81192, 81194, 81197, 81258, 81297, 81499, 81627, 81765, 81767, 81769, 81770, 81773, 81803, 81804, 81823, 81827, 81828, 81829, 81832, 81833, 81834, 81835, 81837, 81838, 81839, 81840, 81841, 81842, 81843, 81844, 81845, 81846, 81853, 81855, 81857, 81858, 81860, 81861, 81862, 81863, 81864, 81865, 81866, 81867, 81868, 81869, 81884, 81885, 81886, 81920, 81921, 81922, 81923, 81924, 81925, 81926, 81927, 81928, 81929, 81930, 81931, 81932, 81933, 81934, 81935, 81936, 81937, 81938, 81939, 81940, 81941, 81942, 81943, 81944, 81945, 81946, 81949, 81957, 81983, 81984, 81985, 81986, 81987, 81988, 81989, 81990, 81996, 81997, 81998, 81999, 82000, 82001, 82003, 82004, 82005, 82016, 82017, 82019, 82021, 82022, 82023, 82024, 82025, 82027, 82028, 82034, 82035, 82036, 82037, 82038, 82039, 82043, 82061, 82062, 82063, 82064, 82065, 82066, 82067, 82068, 82069, 82070, 82071, 82072, 82073, 82074, 82075, 82076, 82077, 82078, 82079, 82080, 82098, 82099, 82100, 82101, 82102, 82103, 82104, 82105, 82106, 82107, 82108, 82109, 82111, 82122, 82123, 82124, 82126, 82127, 82128, 82160, 82161, 82162, 82163, 82164, 82165, 82169, 82172, 82173, 82174, 82175, 82176, 82177, 82178, 82179, 82180, 82181, 82182, 82183, 82184, 82185, 82186, 82192, 82193, 82194, 82195, 82196, 82198, 82199, 82200, 82201, 82202, 82204, 82252, 82253, 82254, 82255, 82256, 82257, 82258, 82259, 82260, 82261, 82262, 82263, 82264, 82265, 82266, 82267, 82268, 82270, 82271, 82272, 82273, 82284, 82285, 82286, 82317, 82318, 82319, 82321, 82322, 82323, 82324, 82325, 82326, 82327, 82328, 82329, 82330, 82342, 82343, 82344, 82346, 82347, 82348, 82349, 82408, 82409, 82411, 82412, 82413, 82414, 82415, 82416, 82417, 82418, 82419, 82421, 82424, 82426, 82430, 82461, 82462, 82463, 82464, 82465, 82466, 82467, 82468, 82469, 82470, 82471, 82472, 82473, 82474, 82475, 82517, 82518, 82519, 82520, 82521, 82522, 82528, 82529, 82530, 82531, 82532, 82533, 82536, 82537, 82538, 82550, 82552, 82553, 82555, 82556, 82557, 82559, 82560, 82561, 82562, 82563, 82564, 82565, 82569, 82570, 82612, 82613, 82614, 82615, 82616, 82617, 82618, 82619, 82621, 82623, 82624, 82625, 82626, 82627, 82628, 82629, 82630, 82631, 82655, 82656, 82657, 82658, 82659, 82660, 82661, 82662, 82663, 82664, 82665, 82666, 82667, 82668, 82669, 82670, 82671, 82672, 82689, 82690, 82773, 82774, 82775, 82776, 82777, 82778, 82779, 82780, 82781, 82782, 82783, 82784, 82785, 82786, 82787, 82788, 82789, 82790, 82791, 82792, 82793, 82794, 82795, 82796, 82797, 82798, 82799, 82800, 82801, 82802, 82803, 82804, 82805, 82806, 82807, 82808, 82809, 82810, 82811, 82812, 82813, 82827, 82828, 82829, 82830, 82852, 82857, 82858, 82859, 82962, 82965, 82966, 82967, 82975, 82986, 82989, 82991, 82998, 83010, 83011, 83012, 83013, 83060, 83061, 83062, 83063, 83064, 83065, 83066, 83067, 83084, 83088, 83089, 83090, 83142, 83144, 83145, 83147
510(k) numbers
["K974034"]
PMA numbers
not on file
Product code
MJK
Firm FEI
1418729
Firm city
Buffalo Grove
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register