Skip to the content

FDA Medical Device Recalls (openFDA)

ICU Medical, Inc.: ICU Medical Double ChemoLock Port Syringe Transfer Set REF: CL-33

Recall number
Z-0342-2026
Recalling firm
ICU Medical, Inc.
Product
ICU Medical Double ChemoLock Port Syringe Transfer Set REF: CL-33
Status
Open, Classified
Initiated
2025-09-08
Posted
2025-10-27
Terminated
not on file
Reason
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
Root cause
Under Investigation by firm
Action
On September 8, 2025, ICU Medical issued an "Urgent Medical Device" Recall notification via E-Mail. ICU Medical ask consignees to take the following actions: 1. Check all inventory locations within your institution for the affected products listed in recall notification and discontinue use. Destroy all affected products following your institution's process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 2. Share this notification with all potential users of the device, to ensure they are aware of this notification. If the devices are used at another location, please ensure this communication is delivered there. 3. Complete and return the attached Customer Response Form to marketaction@mailac.custhelp.com within 10 days of receipt to acknowledge your understanding of this notification. 4. If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to marketaction@mailac.custhelp.com For further inquiries, please contact ICU Medical Customer Service via email: customerservice@icumed.com or call: 1-(800)-258-5361 for additional information or technical assistance. Or contact Field Service Processing via email: marketaction@mailac.custhelp.com for Assistance with Customer Notifications and Response Forms.
Quantity
4000 units
Distribution
Worldwide distribution: US (Nationwide) including states of: AK AL AR AZ CA CO CT DE FL GA IA ID IL KS KY LA MA MD MI MN MS MT NC NE NJ NM NV NY OH OK OR PA SD TN TX VA WA WY and OUS (foreign) to countries of: Canada, France, and Japan.
Lot, serial or model codes
UDI:10840619061729/Lot: 14393155, 14400613, 14376051
510(k) numbers
["K131549"]
PMA numbers
not on file
Product code
ONB
Firm FEI
2025816
Firm city
San Clemente
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register