FDA Medical Device Recalls (openFDA)
Elekta, Inc.: Elekta Synergy XVI X-ray Volume Imaging System To be used as part of radiation therapy treatment process.
- Recall number
- Z-0342-2012
- Recalling firm
- Elekta, Inc.
- Product
- Elekta Synergy XVI X-ray Volume Imaging System To be used as part of radiation therapy treatment process.
- Status
- Terminated
- Initiated
- 2011-10-20
- Posted
- 2011-12-01
- Terminated
- 2012-08-09
- Reason
- A small number of sites have experienced situations with kV generator transformers overheating due to a diode failure on the interface board.
- Root cause
- Nonconforming Material/Component
- Action
- The firm, Elekta, issued a "FIELD CHANGE ORDER MANDATORY ACTION" notice dated October 20, 2011 to its customers to add fuse protection for the kV transformers. The notice describes the product, problem, provides instructions and procedures and actions to be taken. The customers were instructed to follow any instructions or recommendations covered in the notice and to file the document in the Important Notice section of the appropriate User Manual. Elekta will contact the customer if the FCO is applicable to their kV generator. If you have any questions, call 770-300-9725.
- Quantity
- 123
- Distribution
- Nationwide distribution: USA including states of: AL, AK, AZ, CA, CT, FL, GA, IL, IA, KY, MD, MS, MI, MN, MS, MO, MT, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TX, VT, VA, WA, WV, WI, Puerto Rico, & DC.
- Lot, serial or model codes
- G30369, G29103, G34727, G32313, G30367, G29101, G31330, G34401, G32304, G30962, G30610, G30668, G29986, G28427, G28581, G28673, G28430, G28755, G28576, G28574, G28676, G28678, G28811, G28763, G28806, G28863, G28809, G28864, G28936, G29035, G29042, G30720, G28929, G29043, G29102, G28421, G29296, G29299, G29843, G31188, G29832, G29838, G29844, G29840, G29983, G30300, G30027, G29980, G30210, G30213, G30034, G30208, G30309, G30310, G30530, G30533, G30558, G30565, G32986, G30606, G30611, G30961, G30718, G30662, G30717, G31397, G31326, G31186, G31190, G31323, G29842, G34393, G31404, G31512, G31522, G31518, G31515, G31517, G30073, G32359, G32363, G32305, G32364, G32366, G32385, G32425, G32389, G32501, G32680, G32682, G32809, G32814, G34285, G32905, G32820, G32821, G32824, G32904, G32908, G33335, G33339, G33338, G33428, G33433, G34150, G34288, G34151, G34159, G34290, G34348, G34347, G34350, G34397, G34501, G34684, G34502, G34505, G34689, G30974, G34691, G34980, G34983, G34726
- 510(k) numbers
- ["K100115"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 1037831
- Firm city
- Norcross
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28