Skip to the content

FDA Medical Device Recalls (openFDA)

Elekta, Inc.: Elekta Synergy XVI X-ray Volume Imaging System To be used as part of radiation therapy treatment process.

Recall number
Z-0342-2012
Recalling firm
Elekta, Inc.
Product
Elekta Synergy XVI X-ray Volume Imaging System To be used as part of radiation therapy treatment process.
Status
Terminated
Initiated
2011-10-20
Posted
2011-12-01
Terminated
2012-08-09
Reason
A small number of sites have experienced situations with kV generator transformers overheating due to a diode failure on the interface board.
Root cause
Nonconforming Material/Component
Action
The firm, Elekta, issued a "FIELD CHANGE ORDER MANDATORY ACTION" notice dated October 20, 2011 to its customers to add fuse protection for the kV transformers. The notice describes the product, problem, provides instructions and procedures and actions to be taken. The customers were instructed to follow any instructions or recommendations covered in the notice and to file the document in the Important Notice section of the appropriate User Manual. Elekta will contact the customer if the FCO is applicable to their kV generator. If you have any questions, call 770-300-9725.
Quantity
123
Distribution
Nationwide distribution: USA including states of: AL, AK, AZ, CA, CT, FL, GA, IL, IA, KY, MD, MS, MI, MN, MS, MO, MT, NE, NH, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TX, VT, VA, WA, WV, WI, Puerto Rico, & DC.
Lot, serial or model codes
G30369, G29103, G34727, G32313, G30367, G29101, G31330, G34401, G32304, G30962, G30610, G30668, G29986, G28427, G28581, G28673, G28430, G28755, G28576, G28574, G28676, G28678, G28811, G28763, G28806, G28863, G28809, G28864, G28936, G29035, G29042, G30720, G28929, G29043, G29102, G28421, G29296, G29299, G29843, G31188, G29832, G29838, G29844, G29840, G29983, G30300, G30027, G29980, G30210, G30213, G30034, G30208, G30309, G30310, G30530, G30533, G30558, G30565, G32986, G30606, G30611, G30961, G30718, G30662, G30717, G31397, G31326, G31186, G31190, G31323, G29842, G34393, G31404, G31512, G31522, G31518, G31515, G31517, G30073, G32359, G32363, G32305, G32364, G32366, G32385, G32425, G32389, G32501, G32680, G32682, G32809, G32814, G34285, G32905, G32820, G32821, G32824, G32904, G32908, G33335, G33339, G33338, G33428, G33433, G34150, G34288, G34151, G34159, G34290, G34348, G34347, G34350, G34397, G34501, G34684, G34502, G34505, G34689, G30974, G34691, G34980, G34983, G34726
510(k) numbers
["K100115"]
PMA numbers
not on file
Product code
IYE
Firm FEI
1037831
Firm city
Norcross
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register