Skip to the content

FDA Medical Device Recalls (openFDA)

Axya Medical: Axya Bone Anchor 3.0 mm Bioabsorbable with UHMWPE Suture,Catalog Number: CAT1266, Axya Medical, Inc., Beverly, MA 01915

Recall number
Z-0342-2008
Recalling firm
Axya Medical
Product
Axya Bone Anchor 3.0 mm Bioabsorbable with UHMWPE Suture,Catalog Number: CAT1266, Axya Medical, Inc., Beverly, MA 01915
Status
Terminated
Initiated
2007-09-27
Posted
2007-12-08
Terminated
2012-06-21
Reason
Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging
Root cause
Package design/selection
Action
Axya Medical notified customers by letter, Urgent-Product Recall, dated 9/25/07 , via UPS . Users are requested to return product and complete the reply form.
Quantity
111 units
Distribution
Nationwide
Lot, serial or model codes
Lot Numbers: 204677, 204800, 204826 and 204838
510(k) numbers
["K051983","K052491","K060970"]
PMA numbers
not on file
Product code
MBI
Firm FEI
3000718467
Firm city
Beverly
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register