Fresenius Kabi, LLC: The box label states: "ATF 40 Fast Start Kit 9108491...STERILE...FRESENIUS KABI...Fresenius Kabi AG 61345 Bad Hamburg Germany...Federal (USA) Law restricts this device to sale by or on order of a physician". The ATF 40 Fast Start Kit contains one each AT1 Autotransfusion Set, ATS Suction Line, and the ATR40 Autotransfusion Reservoir. ATS is a suction line to aspirate and add anticoagulate to shed blood from the surgical field into an autotransfusion reservoir (ATR). The ATS has a inch. suction port and is connected to a vacuum source. ATR is an autotransfusion reservoir to collect, de-foam, filter, and store blood prior to processing. The ATR reservoirs are offered with either a 120 micron or a 40 micron filter. ATF 40 Autotransfusion set is a disposable set, dedicated component of the Fresenius CATS Continuous Autotransfusion System, which in turn is an Autotransfusion device for the processing of autologous shed blood collected intraoperatively and postoperatively to obtain washed packed red blood cells for reinfusion.
- Recall number
- Z-0341-2011
- Recalling firm
- Fresenius Kabi, LLC
- Product
- The box label states: "ATF 40 Fast Start Kit 9108491...STERILE...FRESENIUS KABI...Fresenius Kabi AG 61345 Bad Hamburg Germany...Federal (USA) Law restricts this device to sale by or on order of a physician". The ATF 40 Fast Start Kit contains one each AT1 Autotransfusion Set, ATS Suction Line, and the ATR40 Autotransfusion Reservoir. ATS is a suction line to aspirate and add anticoagulate to shed blood from the surgical field into an autotransfusion reservoir (ATR). The ATS has a inch. suction port and is connected to a vacuum source. ATR is an autotransfusion reservoir to collect, de-foam, filter, and store blood prior to processing. The ATR reservoirs are offered with either a 120 micron or a 40 micron filter. ATF 40 Autotransfusion set is a disposable set, dedicated component of the Fresenius CATS Continuous Autotransfusion System, which in turn is an Autotransfusion device for the processing of autologous shed blood collected intraoperatively and postoperatively to obtain washed packed red blood cells for reinfusion.
- Status
- Terminated
- Initiated
- 2010-03-29
- Posted
- 2010-11-10
- Terminated
- 2011-06-15
- Reason
- Using AT1 Autotransfusion Set from specific batches in combination with the auto transfusion device CATS Continuous Autotransfusion has led to insufficient suction of shed blood from the collection reservoir. The results in the failure message "alarm blood flow" and the CATS device stops in a secure mode.
- Root cause
- Other
- Action
- Fresenius Kabi sent out a Technical Notification letter dated March 29, 2010, to one distributor located in Ann Arbor, Michigan. The distributor sent the technical notification with a cover letter to each of their customer's that had ordered AT1's, ATF120's and ATF40's over the last year. The letter identified the product, the problem, and the action to be taken. Customers were notified that the autotransfusion prime and / or washing procedure can be continued after alarm blood flow by following Alternative procedures. The Technical Notification informs the distributor about the problem of the AT1 (item #9005101), but also mentions the other two products ATF120 (item #9108501) and ATF40 (item #9108491) which have no known problem. The distributor is advised to contact Fresenius Kabi, LLC for any question regarding the Technical Notification. For questions regarding this recall call (425) 242-2074.
- Quantity
- 17, 444 units distributed in US
- Distribution
- Worldwide Distribution - USA and the countries of Australia,Austria, Brazil, Canada, China, Denmark, France, Germany, Hong Kong, India, Italy, Japan, Malaysia, Netherlands, Phillipines, Portugal, Switzerland, Thailand, UK, and Vietnam.
- Lot, serial or model codes
- Lot numbers: YHT242, YIT012, YIT291, YKT011, YKT142, YKT283, YLT302, ZAT262, ZBT125, ZCT 032, and ZCT 295.
- 510(k) numbers
- ["K960006"]
- PMA numbers
- not on file
- Product code
- CAC
- Firm FEI
- 3007344667
- Firm city
- Redmond
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28